X线管电压对生物组织CT值影响的实验研究 被引量:10

The effects of X-ray tube voltage on tissue CT value
在线阅读 下载PDF
导出
摘要 目的:探讨X线管电压对兔器官组织CT值的影响。方法:在同一台CT机上采用4组X线管电压80、100、120和140kV(头部模式只有前3组管电压,体部模式4组管电压)扫描12只实验兔,在切换管电压时保持其它扫描参数不变。用头颅序列扫描实验兔的头部,用体部序列扫描实验兔的胸部和腹部,测量实验兔的皮质骨、脑组织、肌肉、肝脏、肾脏、脂肪、肺组织的CT值。结果:X线管电压的改变对兔器官组织的CT值影响不一致。脂肪、肝脏、肾脏和皮质骨不同管电压下的CT值差异有统计学意义(P值均<0.05),肺组织、脑组织和肌肉不同管电压下的CT值差异无统计学意义(P值均>0.05)。结论:兔部分器官组织的CT值与X线管电压有一定相关性。临床中可根据组织器官的扫描参数来设定参考CT值。 Objective:To explore the effects on CT values of rabbit organs and tissues caused by X-ray tube voltage variation. Methods: Four groups of X-ray tube voltage (80kV, 100kV, 120kV, 140kV) were used for scanning of 12 rabbits with the other parameters unchanged. The head was scanned with 80,100 and 120kV, and the abdomen was scanned with 80,100,120 and 140kV. The CT values of different tissues were measured and analyzed including bone, brain,muscle,liver, kidney,fat and lung. Results:The fat,liver, kidney, bone were affected significantly by the X-ray tube voltage (P〈0.05). The CT numbers changed remarkably with tube voltage changing. The high density bone,liver and kidney had negative rela tivity to tube voltage, while the fat has positive relativity to tube voltage. The lung, brain and muscle, had no significant effect on X-ray tube voltage (P〉0.05). Conclusions: Some rabbit organs and tissues have different CT values with different X-ray tube voltage and have their own special relativities to the X ray tube voltage. We can set up reference CT values ac-cording to the scan parameters of tissues.
出处 《放射学实践》 2013年第11期1102-1104,共3页 Radiologic Practice
基金 温州科技局资助课题(Y20100344)
关键词 体层摄影术 X线计算机 CT值 组织 Tomography,X-ray computed CT value Tissues
作者简介 彭文献(1977-),男,浙江瑞安人,硕士,高级工程师,主要从事放射成像技术和影像图像处理研究。 通讯作者:夏顺仁,Email:srxia@zju.edu.cn
  • 相关文献

参考文献9

二级参考文献14

  • 1祁振宇,黄劭敏,邓小武.放疗计划CT值的校准检测及其影响因素分析[J].癌症,2006,25(1):110-114. 被引量:53
  • 2Schneider W, Bortfeld T, Schlegel W. Correlation between CT numbers and tissue parameters needed for Monte Carlo simulations of clinic dose distributions. Phys Med Biol,2000,45 (2) :459-478.
  • 3翁学军.蒙特卡罗剂量计算方法的研究.南京:东南大学,2003:60-67.
  • 4MUTIC S, PALTA J R, BUTKER E K, et al. Quality assurance for computed tomography simulators and the computed tomography simulation process: report of the AAPM Radiation Therapy Committee Task Group No. 66 [J]. Med Phys, 2003,30 (10) :2762-2793.
  • 5BENEVENTI S, CHIONNE F,GOBBI G, et al. Quantitative CT tomography for radiotherapy treatment planning: calibration phantom andsources of error [J]. Radiother Oncol, 1995, 37 (supplement): 41.
  • 6LUCA C, ANTONELLA F,FRANCESCA B, et al. Dosimetric impact of computed tomography calibration on a commercial treatment planning system for external radiation therapy [J]. Radiother oncol, 1998,48(3) :335-338.
  • 7METCALFE P. Radiotherapy treatment planning [M]// METCALFE P,KRON T, eds. The physics of radiotherapy X-rays. Oxford : Oxford university press, 2000:293-364.
  • 8MACKIE T R, EL-KHATIB E,BATTISTA J, et al. Lung dose correction for 6 and 15 MV X-rays[J]. Med Phys, 1985,12(3) :327.
  • 9CONSTANTINOU C, HARRINGTON J C, CADIEUX R A, et al. Dosimetry at lung-muscle and lung-bone interfaees using 6 MV and 10 MV X-rays and 12-18 MeV electrons [J].Int J Radiat Oncol Biol Phys, 1993,15(supplement 1 ) : 244.
  • 10CONSTANTINOU C, HARRINGTON J C, DEWERD L A. An electron density phantom for calibrating CT-based planning computers to correct for heterogeneities [J]. Med Phys,1992, 19(2) :325-332.

共引文献91

同被引文献97

引证文献10

二级引证文献32

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部