基于原始数据的迭代重建技术在眼眶CT平扫中的应用 被引量:4

Application of sinogram affirmed iterative reconstruction technique in orbital CT scan
在线阅读 下载PDF
导出
摘要 目的探讨基于原始数据的迭代重建(SAFIRE)技术在眼眶CT平扫中的价值。方法对50例眼部外伤患者行眼眶CT平扫,分别采用用滤过反投影(FBP)和SAFIRE(强度1~5级)技术重建图像;对比分析6种重建图像的平均CT值、噪声、SNR、CNR、图像质量评分及病变检出率。结果 6种重建图像间,同种组织CT值差异无统计学意义(P均>0.05);与FBP图像相比,1~5级SAFIRE图像中视神经噪声分别降低15.69%、27.39%、36.17%、41.75%和53.99%,视神经SNR分别提高18.62%、37.46%、55.70%、69.55%和118.11%,CNR分别提高21.59%、32.90%、56.43%、83.91%和104.39%(P均<0.05);SAFIRE 1级和5级图像的质量评分低于FBP图像,2~4级图像的质量评分高于FBP图像,SAFIRE 3级图像评分最高。结论眼眶CT平扫中应用SAFIRE能有效降低噪声和提高图像质量。 Objective To investigate the value of sinogram affirmed iterative reconstruction (SAFIRE) technique in orbital CT scan. MethOds Totally 50 ocular trauma patients underwent orbital CT scan, and six groups images were recon- structed with FBP and SAFIRE (strength grade 1-5). CT value, noise, SNR, CNR, image quality scoring and lesion detection rate among FBP and SAFIRE images were compared. Results There was no difference among 6 groups reconstructed images on CT value of the same tissue (P〉0.05). Compared with FBP image, noises of optic nerve in SAFIRE 1-5 image decreased by 15.69%, 27.39%, 36.17%, 41.75% and 53.99%, SNR of optic nerve increased by 18.62%, 37. 46%, 55.70%, 69.55% and 118. 11%, while CNR increased by 21.59%, 32.90%, 56.43%, 83.91% and 104.39%, respectively (all P〈0.05). Quality scores of SAFIRE 1 and SAFIRE 5 image were lower, while SAFIRE 2-4 image were higher than that of FBP, and the score of SAFIRE 3 image was the highest. There was no obvious difference among all the images on the rate of lesion detection. Conclusion SAFIRE can effectively decrease the image noise and increase the image quality in orbital CT scanning.
出处 《中国医学影像技术》 CSCD 北大核心 2013年第6期1019-1022,共4页 Chinese Journal of Medical Imaging Technology
关键词 基于原始数据的迭代重建 噪声 眼眶 体层摄影术 X线计算机 Sinogram affirmed iterative reconstruction Noise Orbit Tomography, X-ray computed
作者简介 曾宪春(1979-),男(回族),贵州普安人,在读硕士,主治医师。研究方向:影像诊断。E—mail:zengxianchun04@foxmail.com E-mail:kmhandan@sina.com
  • 相关文献

参考文献12

  • 1Brenner DJ, Hall sEJ. Computed tomography—an increasingsource of radiation exposure. N Engl J Med, 2007, 357 (22):2277-2284.
  • 2Hein E, Rogalla P, Kingebiel R, et al. Computed tomographyspiral CT low-dose CT paranasal sinuses radiation dose eye lensradioprotection. Eur Radiol, 2002,12(7): 1693-1696.
  • 3徐卓东.CARE kV—智能最佳kV扫描技术——降低剂量的同时提高图像质量[J].中国医疗设备,2012,27(2):120-123. 被引量:29
  • 4Price RR, Axel L, Morgan T, et al. Quality assurance methodsand phantoms for magnetic resonance imaging: Report of AAPMnuclear magnetic resonance Task Group No. 1. Med Phys,1990,17(2):287-295.
  • 5Wang G, Yu H, De Man B. An outlook on X-ray CT researchand development. Med Phys, 2008, 35(3) : 1051-1064.
  • 6Kalra MK, Woisetschlager M, Dahlstrom N,et al. Radiationdose reduction with Sinogram Affirmed Iterative Reconstructiontechnique for abdominal computed tomography. J Comput AssistTomogr, 2012, 36(3) : 339-346.
  • 7Thibault JB, Sauer KD, Bouman CA, et al. A three-dimensionalstatistical approach to improved image quality for multislice heli-cal CT. Med Phys, 2007, 34(11) : 4526-4544.
  • 8Moscariello A,Takx RA,Schoepf UJ,et al. Coronary CT angi-ography :Image quality, diagnostic accuracy, and potential forradiation dose reduction using a novel iterative image reconstruc-tion technique-comparison with traditional filtered back projec-tion. Eur Radiol, 2011,21(10):2130-2138.
  • 9Wang R, Schoepf UJ, Wu R, et al. Image quality and radiationdose of low dose coronary CT angiography in obese patients: Si-nogram affirmed iterative reconstruction versus filtered backprojection. Eur J Radiol, 2012,81(11) :3141-3145.
  • 10Winklehner A, Karlo C, Puippe G, et al. Raw data-based itera-tive Reconstruction in body CTA : Evaluation of radiation dosesaving potential. Eur Radiol, 2011, 21(12) : 2521-2526 .

二级参考文献25

  • 1虞崚崴,汤光宇.双源CT冠状动脉血管成像的临床应用[J].中国医学影像技术,2009,25(S1):206-209. 被引量:6
  • 2张兆琪,马晓海.64层螺旋CT冠状动脉成像——无创性冠状动脉检查的新纪元[J].中华放射学杂志,2006,40(8):789-791. 被引量:88
  • 3Wang G, Yu H, De Man B. An outlook on X-ray CT research and development. Med Phys, 2008,35(3) :1051 -1064.
  • 4Leschka S, Scheffel H, Desbiolles L, et al. Image quality and re construction intervals of dual-source CT coronary angiography Recommendations for ECG-pulsing windowing. Invest Radiol 2007,42(8):543- 549.
  • 5Alkadhi H, Scheffel H, Desbiolles L, et al. Dual-source compu ted tomography coronary angiography : Influence o{ obesity, calci um load, and heart rate on diagnostic accuracy. Eur Heart J 2008,29(6) :766-776.
  • 6Ferencik M, Nomura CH, Maurovich-Horvat P, et al. Quantita- tive parameters of image quality in 64-slice computed tomography angiography of the coronary arteries. Eur J Radiol, 2006,57(3) : 373-379.
  • 7Achenbach S, Giesler T, Ropers D, et al. Comparison of image quality in contrast-enhanced coronary-artery visualization by electron beam tomography and retrospectively electrocardiogram- gated multislice spiral computed tomography. Invest Radiol, 2003,38(2) :119- 128.
  • 8Ferencik M, Moselewski F, Ropers D, et al. Quantitative param- eters of image quality in muhidetector spin 1 computed tomographic coronary imaging with submillimeter colimation. Am J Cardiol, 2003,92(11) : 1257-1262.
  • 9Lembcke A, Wiese TH, Schnorr J, et al. Image quality of nonin vasive coronary angiography using multislice spiral computed tomography and electron beam computed tomography: lntraindividual comparison in an animal model. Invest Radiol, 2004, 39 (6) :357-364.
  • 10Pontana F, Pagniez J, Flohr T, et al. Chest computed tomography using iterative reconstruction vs filtered back projection (Part 1): Evaluation of image noise reduction in 32 patients. Eur Radiol, 2011,21 (3) : 627-635.

共引文献67

同被引文献40

  • 1Brenner DJ, Hall EJ. Computed tomography-an increasing source of radiation exposure[J]. N Engl J M,2007,357(22) :2277-2284.
  • 2Hein E,Rogalla P,Kingebiel R,et al. Computed tomography spiral CT low-dose CT paranasal sinuses radiation dose eye lens radio- protection[J]. Eur Radiol, 2002,12(7) : 1693-1696.
  • 3Menzel H, Schihilla H, Teunen D, et al. European guidelines on quality criteria for computed tomography[M]. Luxembourg:Eu- ropean Commission,2000 : 162-163.
  • 4Price RR,Axel L,Morgan T,et al. Quality assurance methods and phantoms for magnetic resonance Imaging: report of AAPM nu- clear magnetic resonance task Group No. 1 [J]. Med Phys, 1990, 17(2) :287-295.
  • 5Pierce DA, Preston DL. Radiation-related cancer risks at low doses among atomic bomb survivors[J]. Radiat Res, 2000,154 ( 2 ) : 178- 186.
  • 6Brenner DJ, Elliston CD. Estimated radiation risks potentially as sociated with full-body CT screening [J]. Radiology, 2004, 232(3):735-738.
  • 7Hundt W,Rust FP,Stabler AM,et al. Dose reduction in multislice computed tomography[J]. J Comput Assist Tomogr, 2005,29 (1) : 140-147.
  • 8Winklehner A, Goetti R, Baumueller S, et al. Automated attenua tion-based tube potential selection for thoracoabdominal compu ted tomography angiography:improved dose effectiveness[J]. In vest Radiol,2011,46(12) :767-773.
  • 9Eller A, May MS, Scharf M, et al. Attenuation-based automatic kilovolt selection in abdominal computed tomography: effects on radiation exposure and image quality[J]. Invest Radiol, 2012,47 (10) :559-565.
  • 10Slovis TL. CT and computed radiography: the pictures are great, but is the radiation dose greater than required[J]. AJR, 2002, 179(1) :39-41.

引证文献4

二级引证文献22

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部