摘要 目的:探讨两种品牌快速血糖仪快速测定末梢血葡萄糖的效果。方法:选择120例住院患者为研究对象,分别采用URIT-86免调码血糖测试系统和强生Nova血糖测试系统测定血糖,比较两种方法测定血糖的差异并与生化分析仪进行比较。结果:两种品牌血糖仪与生化分析仪之间线性回归分析结果均r^2>0.95,URIT-86免调码血糖测试系统在本试验中准确度优于强生Nova血糖测试系统。结论:两种品牌血糖仪检测结果准确可靠,可用于医疗机构及家庭进行血糖监测。URIT-86免调码血糖仪相较强生Nova血糖仪检测结果更优。 Objective:To explore the effect of two kinds of brand fast glucometer in the fast determination of end blood glucose.Methods:120 patients admitted to hospital were served as the research subjects.The blood glucose was measured by URIT-86 code-free blood glucose test system and Johnson Nova blood glucose test system.The differences between the two methods were compared,and also compared them with the biochemistry analyzer.Results:The results of linear regression analysis between the two brands of blood glucose meters and biochemical analyzers were all more than r2>0.95.The accuracy of URIT86 code-free blood glucose test system was better than that of Johnson Nova blood glucose test system.Conclusion:The two kinds brand of glucometer were accurate,may use in medical institute and family.URIT-86 code-free glucometer was better than that of Johnson Nova glucometer.
作者 林敏杰 林芳 李秋平 唐艳华 曹艺 谢文俊 LIN Minjie;LIN Fang;LIQiuping(Guilin Youlite Medical Electronics Co.,Ltd,Guilin 541000;Health Big Data URIT Laboratory,Shenzhen Research Institute,Beijing University of Aeronautics and Astronautics,Shenzhen 518000,China)
出处 《华夏医学》 CAS 2020年第1期132-135,共4页 Acta Medicinae Sinica
关键词 即时检验 快速血糖仪 血糖监测系统 immediate examination fast glucometer blood sugar monitor system
作者简介 通信作者:林芳,教授,Email:linfang@uritest.com.cn。
  • 相关文献

参考文献1

二级参考文献10

  • 1International Organization for Standardization. In vitro diagnostic test systems-Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus [ EB/OL]. (2013- 05-15 ) [ 2016437-10 ]. http://www, iso. org/iso/catalogue _ detail9 csnumber = 26309.
  • 2中华人民共和国国家食品药品监督管理局.体外诊断检验系统自测用血糖监测系统通用技术条件[S/OL].(2006-01-10)[2016-07-10].http://www.sda.gov.cn/ws01/CL059/9410.html.
  • 3中国合格评定国家认可委员会.CNAS-CL38:医学实验室质量和能力认可准则在临床化学检验领域的应用说明[S/OL].(2012-09-13) [ 2016-07-10 ]. https ://www. cnas. org. cn/rkgf/ sysrk/rkyyzz/2014/05/782592, shtml.
  • 4Tang Z, Du X, Louie RF, et al. Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer~ J ]. Am J Clin Patho1,2000,113 ( 1 ) :75-86.
  • 5Ginsberg BH. Factors affecting blood glucose monitoring: sources of errors in measurement [ J]. J Diabetes Sci Technol, 2009, 3 (4) :903-913. DOI: 10.1177/193229680900300438.
  • 6Vote DA, Doar O, Moon RE, et al. Blood glucose meter performance under hyperbaric oxygen conditions [ J ]. Clin Chim Acta ,2001,305 ( 12 ) :81-87.
  • 7Medicines and Heahhcare products Regulatory Agency. Blood glucose meters: point-of-care testing. Advice for heahhcare professionals on the use and management of blood glucose meters [ EB/OL]. (2005-09-14) [ 2016-07-10 ]. https://www, gov. uk/government/publications/blood-glucose-meters-point -of-care- testing.
  • 8Guzel O, Guner El. ISO 15189 accreditation: Requirements for quality and competence of medical laboratories, experience of a laboratory I [ J ]. Clin Biochem, 2009, 42 (4-5) : 274-278. DOI : 10. 1016/j. clinbiochem. 2008.09.011.
  • 9Clinical and Laboratory Standards Institute. CLSI POCT12-A3: Point-of-Care Blood Glucose Testing in Acute and Chronic Care Facilities ; Approved Guideline - Third Edition [ S/OL ]. [ 2016- 07-10]. http://webstore, ansi. org/RecordDetail, aspx? sku = CLSI + POCT12-A3.
  • 10POCT临床应用建议[J].中华检验医学杂志,2012,35(1):10-16. 被引量:42

共引文献76

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部