Influence of reconstruction techniques on PET/CT image quality and quantitative accuracy:a phantom study

原文传递
导出
摘要 Purpose To investigate the influence of different reconstruction techniques on the quantitative accuracy and image quality of PET/CT.Methods The NEMA NU2-2018 image quality phantom was scanned on a GE Discovery Elite PET/CT scanner and the spatial resolution was measured based on NEMA NU2 standard.The same raw data were reconstructed using five different algorithms:ordered subset expectation maximization(OSEM),OSEM with point spread function(PSF)modeling,OSEM with time-of-flight(TOF),OSEM with TOF and PSF,and filtered back-projection(FBP).The recovery coefficient(RC),contrast recovery coefficient(CRC),and contrast-to-noise ratio(CNR)were calculated for the six hot spheres,the percent background variability(PBV)and coefficient of variation(COV)were calculated for the background,and the residual error(RE)was calculated for lung insert in different image slices.Results The incorporation of PSF modeling showed the smallest transverse FWHM and FWTM at both 1 and 10 cm radical offsets.The combination of PSF modeling and TOF improved RC_(mean) and CRC for all spheres and resulted in the highest ratings for the detectability of 10 mm spheres in human observer assessment.PSF modeling played a role in reducing the COV within the background region of interest and increasing the CNR of the spheres,and decreased background noise ratings in human observer assessment.Besides,TOF significantly reduced the RE in lung insert.Neither PSF modeling nor TOF had a significant effect on PBV.Compared to FBP,the OSEM algorithm showed significant advantages in PBV,COV,CNR and RE and human observer ratings of image quality,but worse results for RC_(max),RC_(mean) and CRC.Conclusions The integration of TOF and PSF modeling into the OSEM algorithm achieves improvements in both quantitative accuracy and image quality,providing distinct advantages.PSF modeling improves the spatial resolution and decreases the visual appearance of background noise.The OSEM algorithm shows significantly better image quality than the FBP algorithm but no distinct advantages concerning quantitative accuracy.
出处 《Radiation Detection Technology and Methods》 CSCD 2024年第2期1171-1186,共16页 辐射探测技术与方法(英文)
作者简介 Correspondence author:Jianhua Geng,gengjean@163.com。
  • 相关文献

参考文献6

二级参考文献74

  • 1陈英茂,邵明哲,田嘉禾,尹大一,何义杰,姚树林.正电子发射断层显像中感兴趣区选取及标准摄取值分界点的优化[J].中国医学影像技术,2004,20(6):935-937. 被引量:5
  • 2陈英茂,耿建华,田嘉禾,尹大一,姚树林,陈盛祖.正电子发射断层扫描图像部分容积效应的模拟研究及实验检验[J].原子能科学技术,2004,38(5):467-473. 被引量:6
  • 3耿建华,陈英茂,陈盛祖,田嘉禾,何义杰,乔士珍.符合线路显像与PET显像中SUV的比较研究[J].中华核医学杂志,2004,24(5):308-309. 被引量:11
  • 4Lenders K,Perani D,Lammertsma A,et al.Cerebral blood flow,blood volume and oxygen utilization[J].Normal values and effect of age.Brain,1990,113:27-47.
  • 5Marchal G,Rioux P,Petit-Taboue M-C,et al.Regional cerebral oxygen consumption,blood flow,and blood volume in healthy human aging[J].Arch Neurol,1992,49:1013-1020.
  • 6Meltzer CC,Cantwell MN,Greer PJ,et al.Does cerebral blood flow decline in healthy aging? A PET study with partial-volume correction[J].Nucl Med,2000,41:1842-1848.
  • 7Giovacchini G,Lerner A,Toczek MT,et al.Brain incorporation of 11C-arachidonic acid,boold volume,and blood flow in healthy aging:a study with partial-volume correction[J].Nucl Med,2004,45:1471-1479.
  • 8Frouin V,Comtat C,Reilhac A,et al.Correction of Partial Volume Effect for PET Striatal Imaging:Fast Implementation and Study of Robusmess[J].Nucl Med,2002,43:1715-1726.
  • 9Bencherif B,Stumpf MJ,Links JM,et al.Application of MRI-based partial-volume correction to the analysis of PET images of u-opioid receptors using statistical parametric mapping[J].Nucl Med,2004,45:402-408.
  • 10Quarantelli M,Berkouk K,Prinster A,et al.Integrated software for the analysis of brain PET/SPECT studies with partial-volume-dffect correction[J].Nucl Med,2004,45:192-201.

共引文献47

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部