摘要
采用硝酸和稀硫酸溶解样品,选择Ti 323.966 nm为分析谱线,建立了电感耦合等离子体发射光谱法(ICP-OES)测定银铜钛合金中钛含量的分析方法。实验结果表明,钛的质量浓度在0~80μg/mL范围内与发射强度呈线性相关,校准曲线的线性回归方程为y=83.132 x+21.113,相关系数r=0.999973。该方法中钛的检出限为0.071μg/mL。本实验方法用于银铜钛合金中钛含量的分析,检测结果的相对标准偏差RSD﹤1%(n=10),加标回收率大于99%。与传统化学分析方法进行比对,测定结果一致。
The samples were dissolved by nitric acid and dilute sulfuric acid,and Ti 323.966 nm was selected as the analytical spectral line.The experimental results showed that there was a linear correlation between the emission intensity and the mass concentration of titanium in the range of 0—80μg/mL,the linear regression equation of the calibration curve was y=83.132x+21.113,and the correlation coefficient was 0.999973.The detection limit of titanium was 0.071μg/mL.The relative standard deviation(RSD)was less than 1%(n=10),and the recovery was more than 99%.Compared with the traditional chemical analysis method,the determination results of the two methods were consistent.
作者
张玲玲
金莹
王蒙蒙
Zhang Lingling;Jin Ying;Wang Mengmeng(Zhejiang Metallurgical Research Institute Co.,Ltd.,Hangzhou 310030,China;Zhejiang Asia General Soldering&Brazing Material Co.,Ltd.,Hangzhou 310030,China;Zhejiang Key Laboratory of Brazing Materials and Technology,Hangzhou 310030,China)
出处
《分析仪器》
CAS
2021年第2期134-136,共3页
Analytical Instrumentation
基金
浙江省重点研发计划(2019C01089)。
关键词
湿法消解
电感耦合等离子体发射光谱法
银铜钛合金
钛
Wet digestion
Inductively coupled plasma atomic emission spectrometry(ICP-AES)
Silver-copper-titanium alloy
Ti
作者简介
张玲玲,女,1987年出生,工程师,研究方向为钎焊材料化学分析技术,E-mail:13626712603@163.com。