期刊文献+

基于CFD技术的无叶片离心血泵内部流动特性分析研究 被引量:6

Research on Flow Characteristics in a Non-blade Centrifugal Blood Pump Based on CFD Technology
原文传递
导出
摘要 血泵的血栓和溶血问题是血泵发展至今一直要研究的重要课题。众多研究成果表明,血泵内血液的复杂流动现象和机械运动产生的高剪切力是造成血栓和溶血问题的主要原因。我们通过与上海儿童医学中心合作,应用计算流体动力学(CFD)计算分析软件FLUENT对该院临床中使用的一种无叶片离心式血泵内部流场特性进行分析研究,结果表明该离心泵内部流场分布较为均匀,剪切力分布基本处于红细胞破碎的临界状态之下,血栓和溶血程度均较轻,有良好的应用前景。 The problem of thrombus and hemolysis in blood pump has always been an important topic to study in the development of the blood pump.Numbers of research results show that it is the complicated flow and the high shear stress of the mechanical movement that result in the thrombus and hemolysis.In this study,with the cooperation of Shanghai Children's Medical Center,we have used computational fluid dynamics(CFD) commercial software FLUENT to compute and analyze the flow characteristics in a non-blade centrifugal blood pump.The results figure out that this pump has a reasonable flow distribution and the shear stress distribution is under the critical broken state of red blood cell;meanwhile,there is less thrombus and hemolysis in this pump.So it is in the foreground for clinical use.
机构地区 上海理工大学
出处 《生物医学工程学杂志》 EI CAS CSCD 北大核心 2010年第5期1133-1137,共5页 Journal of Biomedical Engineering
关键词 血泵 血栓 溶血 剪切力 FLUENT Blood pump; Thrombus; Hemolysis; Shear stress; FLUENT;
作者简介 通讯作者。E-mail:y_z_cheng@yahoo.com.cn
  • 相关文献

参考文献5

二级参考文献32

共引文献1259

同被引文献46

  • 1王芳群,钱坤喜.CFD在人工器官设计中的应用[J].北京生物医学工程,2005,24(1):60-63. 被引量:2
  • 2王芳群,封志刚,曾培,茹伟民,袁海宇,钱坤喜.基于CFD的离心式人工心脏泵的溶血预测方法[J].中国生物医学工程学报,2006,25(3):338-341. 被引量:8
  • 3姜华,席光.心脏泵的研究进展[J].医疗卫生装备,2006,27(8):33-35. 被引量:12
  • 4李军,孙皓,李国君,丰镇平.小型离心血液泵内流动特性的数值研究[J].热科学与技术,2007,6(3):235-240. 被引量:3
  • 5Untaroiu A, Throckmorton A L, Patel S M, et al. Nu- merical and experimental analysis of an axial flow left ventricular assist device: the influence of the diffuser on overall pump performance [ J ]. Artif Organs, 2005,29 (7) :581-591.
  • 6Throckmorton A L, Kapadia J Y, Cb.opski S G, et al. Nu- merical hydraulic and hemolytic evaluation of an intravas- cular axial flow blood pump to mechanically support fontan patients[ J]. Ann Biomed Eng, 2011,39( 1 ) :324-336.
  • 7Zhang J, Gellman B, Koert A, et al. Computational and experimental evaluation of the fluid dynamics and hemo- compatibility of the centrimag blood pump [ J ~. Artif Or- gans, 2006,30( 3 ) : 168-177.
  • 8Bludszuweit C. Model for a General Mechanical Blood Damage Prediction [ J ]. Artif Organs, 1995,19 ( 7 ) : 583- 589.
  • 9Song XW,Wood HG.基于计算流体力学的第四代连续流动式心室辅助装置研究[J].美国机械师工程学会传热学会刊,2004,126(4):180-187.
  • 10陶文铨.计算流体力学[M].北京:清华大学出版社,1984.