期刊文献+

两种全自动血液培养系统用于检测脐带血微生物污染的比较 被引量:3

Comparison of two automated blood culture systems in excluding microbial contamination of cord blood
在线阅读 下载PDF
导出
摘要 BacT/ALERT3D240是法国梅里埃公司在BacT/ALERT240基础上改良的具有世界领先水平的新一代血液培养系统,为了解改进后的优势,对我库用新系统检测1268份脐带血和用老系统检测1794份脐带血的结果进行比较分析,以传代培养及涂片镜检为金标准,二者的检测灵敏度、特异性、准确性分别为94.29%、100%、99.84%和90.48%、100%、99.78%,阳性检出时间分别为1.95h至6.0d和2.00h至6.5d。此外,新系统在硬件和软件方面具有明显优势,实用性更强。 BacT/ALERT 3D 240 is a new automated blood culture system improved from BacT/ALERT 240 (bi- oMerieux, France). The two systems were compared on detecting microbial contamination of cord blood in our bank. The sensitivity, specificity, and accuracy were 94. 29%, 100%, and 99. 84% when analyzing 1268 samples with the new system, and they were 90.48% , 100%, 99.78% in evaluation of 1794 samples with the old system. The durations needed for a positive detection were 1.95 h to 6.00 d with the new system and 2.00 h to 6.50 d with the old system. In addition, the new system held the advantages on some functions of hardware and software, which made it more convenient for practical use.
出处 《现代仪器》 2007年第4期45-47,共3页 Modern Instruments
关键词 血液培养系统 评价研究 脐带血 Blood culture system Evaluation study Cord blood
  • 相关文献

参考文献7

二级参考文献16

  • 1Gluckman E,Broxmeyer HE,Averbach AD et al. Hematopoietic reconstitution in a patient with Fanconi' s anemia by means of umbilical -cord blood from an HLA- identical sibling. N Engl J Med 1989,321:1174 - 1178.
  • 2Gluckman E. Current status of umbilical cord blood hematopoietic stem cell transplantation. Exp Hematol 20(10 Nov; 28 ( 11 ) :1197 - 1205.
  • 3Rocha V,Wagner JE,Sobocinski KA,et al. Graft-versus-host disease in children who have received a cord - blood or bone marrow transplant from an HLA - identical sibling. N Engl JMed 2000,342:1846 - 1854.
  • 4Rocha V,Cornish J,Sievers EL,et al. Comparison of outcomes of unrelated bone marrow and umbilical cord blood transplants in children with acute leukemia. Blood 2001,97:2962 -2971.
  • 5Gtuckman E. Ematopoetic stem - cell transplants using umbilical- cord blood. N Engl J Med 2001,24(344 ) : 1860 - 1861.
  • 6Broxmeyer HE,Stout EF,Hangoc Get al. High-efficiency recovery of functional hematopoietic progenitor and stem cells from human cord blood cryopreserved for years. Proc Natl Acad Sci USA 2003. 100(2) :645 -650.
  • 7Rubinstein P, Dobrila L, Rosenfield RE et al. Processing and cryopreservation of placental/umbilical cord blood for unrelated bone marrow reconstitution. Proc Natl Acad Sci USA. 1995,92(22) :10119 -22.
  • 8BioArchive System Operator S Manual. ThermoGenesis Corp.1998, First Issue.
  • 9王霞,汪成孝.脐血移植后GVHD的研究进展[J].中国输血杂志,1997,10(3):153-155. 被引量:3
  • 10张秀珍,宣天芝,胡云建.全自动血培养系统临床应用评价[J].中华医学检验杂志,1998,21(5):297-299. 被引量:59