期刊文献+

医用X线机球管灯丝加热方式的改进 被引量:2

Improvement of Heating Mode of Tube Filament in Medical X-ray Machine
在线阅读 下载PDF
导出
摘要 本文围绕着屡烧X线球管灯丝的4台X线机,通过研究分析引起故障的原因,对设备硬件与软件在设计中存在的问题,进行了完善和改进,成功地对故障机器进行了修复,有效地延长了球管的使用寿命,取得较好的效果。 This article performed a reaearch by revolving around the X-ray tube filament burning repeatedly four X-ray machine.By studying and analyzing the causes of the fault,the problems of the equipment hardware and software in the design were improved and improved,the fault machine was successfully repaired,the service life of the bulb tube was effectively extended,and better results were obtained.
作者 于乃群 汪文标 YU Naiqun;WANG Wenbiao(Department of Equipment, Xiaoshan TCM Hospital of Hangzhou, Hangzhou Zhejiang 311201, China;;Department of Equipment, Xiaoshan Hospital of Zhejiang, Hangzhou Zhejiang 311201, China)
出处 《中国医疗设备》 2018年第12期100-102,133,共4页 China Medical Devices
关键词 X线球管 X线球管灯丝 灯丝加热电路 双板DR X-ray tube X-ray tube filament filament heating circuit double plate DR
作者简介 于乃群,邮箱:hzxsyu@126.com.
  • 相关文献

参考文献11

二级参考文献58

  • 1Mohammad Mehdi NASSERI.Monte-Carlo simulation for determining SNR and DQE of linear array plastic scintillating fiber[J].Nuclear Science and Techniques,2004,15(5):304-307. 被引量:3
  • 2陈玉勤.DR图像的质量控制与保证[J].医疗卫生装备,2005,26(2):42-42. 被引量:19
  • 3白玫,刘旭,彭明辰.数字X射线成像设备DQE的测试[J].医疗设备信息,2007,22(4):1-4. 被引量:13
  • 4白智鹏.医用X射线机原理、构造与维修[M].北京:中国医药科技出版社,1997.
  • 5Chotas HG, Floyd CE, Ravin CE. Technical evalution of a digital chest radiography system that uses a selenium detector[J]. Radiology, 1995, 195: 264-270.
  • 6Launders JH, Cowen AR, Bury RF,et al Towards image quality,beam energy and effective dose optimisation in digitalthoracic radiography[J].Eur Radi01,2001,11: 870-875.
  • 7Schuncke A, Neitzel U. Retrospective patient dose analysis of adigital radiography system in routine clinical use[J].P,adiat Prot Dosimetry,2005, 114:131-134.
  • 8IEC 62220-1.Medical electrical equipment- Characteristics of digital X-ray imaging devices-Part 1.Determination of the detective quantum efficiency[S]. IEC, 2003-10-01.
  • 9国家食品药品监督管理局.YYT0741-2009数字化医用X射线摄影系统专用技术标准[S].国家食品药品监督管理局,2009-11-15.
  • 10Samei E,Flynn MJ,Reimann DA.A method for measuring the presampled MTF of digital radiographic systems using an edge test device[J].Med Phys, 1998,25(1): 102-113.