摘要
目的:探讨CoaguChek XS检测仪在下肢深静脉血栓形成疾病中监测国际标准化比值(international normalized ratio,INR)的准确性。方法:2012年7月-2015年1月在中山市人民医院住院的95例下肢深静脉血栓形成患者,应用床旁快速CoaguChek XS检测仪与我院检验科中心实验室CA7000凝血仪配对检测189组抗凝剂治疗患者的INR。结果:分成三组:1≤INR<2组(n=162)、2≤INR<3组(n=15)及3≤INR组(n=12)。比较两者检测INR的差异及相关性。结果:CoaguChekXS检测仪与CA7000凝血仪检测患者的INR在1≤INR<2,2≤INR<3,3≤INR相关性良好,差异无统计学意义(P>0.05)。在3≤INR组统计有差异(P<0.05)。结论:CoaguChek XS检测仪能够满足下肢深静脉血栓形成患者1≤INR<3的INR监测。3≤INR患者出血风险大,CoaguChek XS检测仪监测INR有差异,需应用CA7000凝血仪复查INR。
Objective:To evaluate the accuracy of the CoaguChek XS system in lower extremity deep venous thrombosis.Methods:95cases were treated in our hospital,International normalized ratio(INR)of patients received anticoagulant,therapy were measured bythe CoaguChek XS system and the CA7000system.INR results was divided into three groups:1≤INR<2(n=162),2≤INR<3(n=15),3≤INR(n=12).The relationship and bias of the INR results tested were analyzed.Results:Correlation analysis indicated that there was asignificant correlation of INR tested by the CoaguChek XS system with those by the CA7000system in groups:1≤INR<2,2≤INR<3.Two systems have differece in3≤INR group.Conclusions:The CoaguChek XS system could be used to monitor the INR(1≤INR<3)inlower extremity deep venous thrombosis.If3≤INR,we need use the CA7000system test INR.
出处
《中医临床研究》
2017年第12期16-17,共2页
Clinical Journal Of Chinese Medicine