迭代重建在膝关节置换术后X线数字化断层融合检查中的应用价值 被引量:6

Application of the Iterative reconstruction technique in digital tomosynthesis examination after knee joint replacement
在线阅读 下载PDF
导出
摘要 目的:探讨X线数字化断层融合(DTS)运用迭代重建技术(IR),降低膝关节图像金属植入物伪影的可行性和临床应用价值。方法:79例人工膝关节置换术后患者行DTS扫描,分别采用滤波反投影重建(FBP)技术和迭代重建(IR)技术对原始图像进行重建。从图像清晰度、假体金属植入物与周围骨相连结构的显示度、金属伪影的多少等方面对两组图像进行分析并评价。根据骨科临床对人工膝关节的分区,在后处理工作站上测量A1~A7区域内假体与骨之间硬化束金属伪影的长度。结果:对两种重建技术所得到的图像进行主观评价,IR组优片率为88.6%,FBP组优片率为62%,IR组优片率高于FBP组,差异有统计学意义(P〈0.05)。IR组重建图像在A1、A3~A7区域无金属伪影,仅3例图像A2区域见少许金属伪影,A7区域关节间隙内聚氨酸脂软垫显示清晰;FBP组人工假体在A1~A7区域均见金属伪影,伪影长度为0.5~2.6mm。FBP组图像伪影明显多于IR组。结论:有金属植入物的人工膝关节置换术后患者行DTS检查,运用IR技术可明显减少金属植入物伪影,图像质量明显改善,在术后随访中具有较高的临床应用价值。 Objective:To study the feasibility and clinical application of using iterative reconstruction technique(IR)in the digital radiographic tomosynthesis(DTS)to reduce the artifact by metallic implant in the knee-joint images.Methods:79patients received knee joint replacement were recruited and underwent DTS scan.The original data were reconstructed by filtered back-projection and iterative reconstruction(IR)technique respectively.These two groups of images were analyzed and scored according to the distinctness of image,the visibility of the connection structure between metallic implants and the surrounding bones and metallic artifacts.The length of beam hardening artifacts between the metallic prosthesis and bones was measured in the fields of A1~A7at the post processing workstation according to the zoning of the artificial knee joint by orthopedics clinical application.Results:The images obtained by the two reconstruction techniques were subjectively evaluated.The satisfactory rate of IR group images was 88.6%,which was higher than that of FBP group(62%),with significant statistic difference(P〈0.05).There was no metallic artifacts in the image of IR group in the field of A1,A3~A7,only few metallic artifacts in the field of A2 were showed in 3cases.In the field of A7,images of polyurethane resin soft pad could be revealed clearly in the joint space.The metallic artifacts of artificial prosthesis of FBP group were found in the field of A1~A7,the length of the artifacts were 0.5~2.6mm.The metallic artifacts in the images of FBP group were obviously more than that of IR group.Conclusion:The artifacts of metallic implant could be reduced by IR technique in patients with artificial knee joint replacement examined by DTS.The image quality was obviously improved,which was of great clinical value in the follow up after knee joint replacement.
出处 《放射学实践》 北大核心 2016年第5期420-423,共4页 Radiologic Practice
关键词 全膝关节置换术 人工膝关节 数字化断层融合 迭代重建 放射摄影术 Total knee arthroplasty Knee prosthesis Digital tomosynthesis Iterative reconstruction Radiography
作者简介 张雪琴(1976-),女,四川人,技师,主要从事影像技术工作。 通讯作者:李真林,E-mail:lzlcdol@126.com
  • 相关文献

参考文献11

二级参考文献62

共引文献106

同被引文献47

引证文献6

二级引证文献25

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部