QC8500型流动注射仪测定水中挥发酚的研究 被引量:3

Determination of Volatile Phenol in Water by QC8500 Flow Injection Analysis
在线阅读 下载PDF
导出
摘要 建立流动注射分析水中挥发酚的方法。实验结果表明,该方法线性范围为0~200μg/L,线性相关系数r为0.999 9;检出限为0.14μg/L;实际水样的加标回收率为93%~102%,精密度和准确度好。本方法与国标法(HJ 503-2009)进行对比,测定结果无明显差异,同时具有分析速度快、灵敏度高、重现性好、自动化程度高、过程安全可靠等优点,可用于水中挥发酚的定量分析。 The method for determination of volatile phenol in water by flow injection analysis was established. The liner range was 0 ~200 μg/L, the correlation coefficient was 0.999 9 and the detection limit was 0.14 μg/L. The recovery range was 93%~102%.The method has good precision and accuracy, consist with GB standard(HJ 503-2009), rapid, sensitive, safe with good accuracy and a high degree of automation, and can be suit to the determination of volatile phenol in water.
出处 《中国环境管理干部学院学报》 CAS 2015年第4期60-62,共3页 Journal of Environmental Management College of China
关键词 流动注射分析 挥发酚 分光光度 flow injection analysis water samples volatile phenol spectrophotometry
作者简介 韩荣荣(1984-),女,河北唐山人,毕业于河北科技大学环境科学专业,助理工程师,主要从事实验室化学及仪器分析工作。