摘要 目的:探讨X线数字断层融合技术在骨骼影像诊断中的应用价值。方法选取我院骨伤科2011年8月~2013年2月收治的经X线普通摄片检查未能确定为骨折的68例患者,对其行数字断层融合技术与CT检查。结果数字断层融合技术检查出隐匿性骨折患者66例(97.05%),而CT检查出隐匿性骨折患者56例(82.35%),两种诊断方法的隐匿性骨折检出率比较,差异有统计学意义(P<0.05)。数字断层融合技术检查的图像优秀率为94.11%,CT检查的图像优秀率为85.29%,两组优秀率比较,差异有统计学意义(P<0.05)。结论数字断层融合技术诊断准确率高、图像清晰、辐射剂量低,值得推广。 Objective To investigate the application value of digital tomosynthesis (DTS) in skeletal imaging diagnostics.Methods 68 patients who could not be identified as fractures with general radiography in our hospital from August 2011 to February 2013 were selected. Then digital tomosynthesis and CT examination were employed to these patients.Results 66 patients (97.05%) were determined as occult fractures by DTS while 56 patients (82.35%) were determined as occult fractures by CT. There were signiifcant differences between the two diagnostic methods on fracture decetion rate (P〈0.05). The proportion of high-quality images obtained by DTS was 94.11% while that of high-quality images obtained by CT examination was 85.29%. There were signiifcant differences between the two methods on the proportion of high-quality images (P〈0.05).Conclusion DTS with high diagnostic accuracy, clear images and low radiation dose is worth to be promoted.
出处 《中国医疗设备》 2014年第7期166-168,共3页 China Medical Devices
关键词 数字断层融合技术 隐匿性骨折 CT机 辐射剂量 digital tomosynthesis obscure bone fracture CT machine radiation dose
作者简介 徐晶,医学在读硕士,主管技师。作者邮箱:xujingsdjn@163.com 通讯作者:邓大平,研究员,硕士生导师。
  • 相关文献

参考文献12

  • 1Sarka V,Shi C,Rassiah-Szegedi P,et al.The effect of a limited number of projections and reconstruction algorithms on the image quality of megavoltage digital tomosynthesis[J].J Appl Clin Med Phys,2009,10(3) :2970.
  • 2James TD,McAdams HP,SongJW,et al.Digital tomosynthesis of the chest for lung nodule detection: interim sensitivity results from an ongoing NIH-sponsored trial[J].Med Phys,2008,35(6):2554-2557.
  • 3Gomi T,Hirano H,Umeda T.Evaluation of the X-ray digital linear tomosynthesis reconstruction processing method for metal artifact reduction[J].Comput Med Imaging Graph,2009,33(4):267-274.
  • 4Mestrovic A,Nichol A,Clark BG,et al.Integration of on-line imaging, plan adaptation and radiation delivery: proof of concept using digital tomosynthesis[J].Phys Med Biol,2009,54(12):3803-3819.
  • 5田军,宋少娟.数字化体层摄影与诊断图谱[M].北京:人民卫生出版社,2012:5-6.
  • 6王巍,刘传亚,卢传友,秦维昌.数字合成X线体层成像原理与研究进展[J].医学影像学杂志,2005,15(7):606-609. 被引量:29
  • 7Godfrey DJ,Rader A,Dobbins JT.Practical strategies for the clinical implementation of matrix inversion tomosynthesis (MITS) [J].Proc SPIE,2003,50(30):379-390.
  • 8Maravilla KR,Murry RC,Deck M,et al.Clinical application of digital tomosynthesis:a preliminary report[J].AJNR Am J Neuroradiol,1983,4(3):277-280.
  • 9Grant DG.Tomosynthesis:a three-dimensional radiographic imaging technique[J].IEEE trans Biomed Eng,1972,19(1):20-28.
  • 10Webber RL,Underhill HA,Hemler PF,et al.Nonlinear algorithm for taskspecitic tomosynthetic image reconstruction[J] .Proc SPIE,1999,3695:258-265.

二级参考文献23

  • 1王巍,刘传亚,卢传友,秦维昌.数字合成X线体层成像原理与研究进展[J].医学影像学杂志,2005,15(7):606-609. 被引量:29
  • 2黄兰珠.非定型硒平板探测器X线摄影应用价值[J].医疗设备信息,2006,21(5):50-51. 被引量:6
  • 3Badea C, Kolitsi Z,Pallikarakis N,et al.A wavelet-based method for removal of out-of-plane structures in digital tomosynthesis[J].Comput Med Imaging Graph,1998,22:309-315.
  • 4Badea C, Kolitsi Z,Pallikarakis N,et al.Image quality in extended arc filtered digital tomosynthesis[J].Acta Radiol,2001,42: 244-248.
  • 5Chakraborty D P, Yester M V, Barnes G T,et al.Selfmasking subtraction tomosynthesis[J].Radiology,1984,150:225-229.
  • 6Dobbins JT,Godfrey DJ. Digital X-ray tomosynthesis:current state of the art and clinical potential[J].Phys Med Biol,2003,48:65-106.
  • 7Dobbins JT.Matrix inversion tomosynthesis improvements in longitudinal x-ray slice imaging[J].Proc SPIE,1990,903:204.
  • 8Dobbins JT,Webber R L,Hames SM.Tomosynthesis for improved pulmonary nodule detection (abstract)[J].Radiology,1998,209:280.
  • 9Duryea J, Dobbins JT III.Application of digital tomosynthesis to hand radiography for arthritis assessment[J]. Proc SPIE,2001,4320:688-695.
  • 10Duryea J, Dobbins JT,Lynch J A. Digital tomosynthesis of hand joints for arthritis assessment[J]. Med Phys,2003,30:325-333.

共引文献69

同被引文献78

  • 1吕玉琦,庞俊.测量桡骨远端骨折骨痂生长的微机图像分析系统[J].中华放射学杂志,1993,27(7):472-475. 被引量:7
  • 2张俊成,赵相胜,张茂根,高建华.直接数字化X线摄影在IVP中的应用[J].实用医技杂志,2005,12(02B):423-425. 被引量:4
  • 3赵怀志,郝华,卢明书,葛长官,王兴国.X光片骨痂定量探讨与应用[J].中国中医骨伤科,1995,3(4):8-9. 被引量:25
  • 4Ren L, Zhang J, Thongphiew D, et d. A novel digital tomosynth- esis(DTS) reconstruction method using a deformation field map [J]. Med Phys, 2008, 35(7): 3110-3115.
  • 5Yan H, Godfrey D J, Yin FF. Fast reconstrution of digital tomosy- nthesis using on board images [J]. Med Phys, 2008, 35(5): 2162- 2169.
  • 6Mermuys K,De Geeter F,Bacher K,et al.Digital tomosynthesis in the detection of urolithiasis: Diagnostic performance and dosimetry compared with digital radiography with MDCT as the reference standard[J].AJR Am J Roentgenol,2010,195 (1) :161-167.
  • 7Remy-Jardin M,Remy J,Giraud F,et al.Pulmonary nodules: detection with thick-section spiral CT versus conventional CT [J].Radiology, 1993,187 (2):513 - 520.
  • 8Rubin GD,Lyo JK,Paik DS,et al.Pulmonary nodules on multi- detector row CT scans:performance comparison of radiologists and computer-aided detection[J].Radiology,2005,234(1): 274-283.
  • 9Gomi T,Nakajima M,Fujiwara H,et al.Comparison between chest digital tomosynthesis and CT as ascreening method to detect artificial pulmonary nodules:a phantom study[J].Br J Radiol,2012,85(1017):e622-629.
  • 10Quaia E,Baratella E,Cernic S,et al.Analysis of the impact of digital tomosynthesis on the radiological investigation of patients with suspected pulmonary lesions on chest radiography[J]. Eur Radiol,2012,22(9):1912-1922.

引证文献9

二级引证文献24

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部