合肥光源钮扣型束流位置检测器的离线标定 被引量:2

Calibration of Button Type Beam Position Monitor for HLSⅡ
在线阅读 下载PDF
导出
摘要 为满足合肥光源重大维修改造项目对储存环束流位置测量的高精度要求,对钮扣束流位置检测器(BPM)电极进行了重新设计,本文对新设计的BPM进行特征阻抗、分布电容和标定系数等参数离线标定。采用时域反射计(TDR)测量其阻抗特性和分布电容,测量结果表明,50Ω标准阻抗测试电缆与BPM Feedthrough连接处的阻抗一致性较好,说明其特性符合设计的50Ω阻抗,计算得到分布电容约为2.9pF,满足设计要求。通过天线法进行标定,分别采用差比和法与对数比法得到直观的mapping图,计算得到BPM的水平和垂直方向的灵敏度,差比和法与对数比法获得的灵敏度相对偏差在垂直方向小于0.5%,水平方向小于2.8%,离线标定获得的各项参数均满足要求。 According to the high precision requirements of the Hefei Light Source major maintenance and reconstruction project , a new button type beam position monitor (BPM) was designed for HLS Ⅱ electron storage ring .The characteristic impedance and distributed capacitance of button pickup were measured by time-domain reflectometer (TDR) ,and scaling coefficient of the new BPM was determined through antenna method .The TDR test results show that the impedance of pickup is 50 Ω which matches the cable impedance ,and the capacitance is about 2.9 pF w hich meets design requirements . Furthermore , the intuitive mapping diagram and calibration factor of beam position were obtained by antenna scanning .Compared with the design value ,the sensitivity deviation of new BPM is less than 0.5% in the vertical direction and less than 2.8% in the horizontal direction .
出处 《原子能科学技术》 EI CAS CSCD 北大核心 2014年第3期541-544,共4页 Atomic Energy Science and Technology
基金 国家自然科学基金资助项目(11105141 11175173) 中央高校基本科研业务费专项资金资助项目(WK2310000015)
关键词 钮扣电极 束流位置检测器 BPM标定 对数比处理 灵敏度 mapping图 button pickup beam position monitor BPM calibration log-ratio processing sensitivity mapping diagram
作者简介 作者简介:杨永良(1980-),男,湖南沅江人,副研究员,从事先进束流诊断技术研究 通信作者:孙葆根,E-mail:bgsun@ustc.edu.cn
  • 相关文献

参考文献7

二级参考文献29

  • 1李吉浩,孙葆根,何多慧,卢平,王筠华,曹涌,郑普.用于合肥光源200MeV直线加速器束流位置测量的信号处理系统[J].强激光与粒子束,2005,17(9):1434-1436. 被引量:4
  • 2LI Ji-Hao SUN Bao-Gen HE Duo-Hui LU Ping WANG Yong WANG Jun-Hua JIN Kai CAO Yong ZHENG Pu.Strip line beam position monitor for HLS LINAC[J].Nuclear Science and Techniques,2005,16(5):257-259. 被引量:4
  • 3赵敬霞,曹建社,裴士伦,叶强,赵晓岩.多条形电极束流能散度探测器的张角优化[J].强激光与粒子束,2006,18(8):1355-1358. 被引量:3
  • 4李吉浩,孙葆根,何多慧,卢平,王筠华,李鹏.HLS直线加速器条带束流位置检测器基于对数比方法的标定[J].原子能科学技术,2007,41(3):339-342. 被引量:5
  • 5RUSSELL J S.Priliminary results of rms emittance measurements performed on the sub-picosecond accelerator using beam position monitors[C]//Proceedings of PAC 1997.[S.l.]:[s.n.],2997:177-179.
  • 6SUWADA T,SATOH M,FURUKAWA K.Nondestructive beam energy-spread monitor using multi-strip-line electrodes[J].Phy Rev,2003,6:032801.
  • 7LI P,SUN B G.New methods of measuring emittance using beam position monitors[C]//Proceedings of 2007 IEEE Nuclear Science Symposium Conference.[S.l.]:[s.n.],2007:1675-1678.
  • 8Libera brilliance single pass user manual 2.02[OL].[2009-10-22].http://www.i-tech.si.
  • 9OLMOS A,G(U)NZEL T,P(E)REZ F,et al.BPM design for the ALBA synchrotron[C]//Proceedings of EPAC.Edinburgh,Scotland:[s.n.],2006:1190-1192.
  • 10KUBE G,WERNER M.Signal level calculation for the PETRA III beam position monitor system[C]//Proceedings of DIPAC.Venice,Italy:[s.n.],2007:105-107.

共引文献9

同被引文献6

引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部