期刊文献+

皮秒拍瓦激光的参数测量系统可靠性分析 被引量:5

Analysis on Reliability of Diagnostics System for Picosecond Petawatt Laser
原文传递
导出
摘要 皮秒参数测量系统用于提供皮秒拍瓦激光系统的各项状态参数,协助激光系统达到预期的技术指标。针对皮秒拍瓦激光系统的技术指标,皮秒参数测量系统将提供压缩脉冲的能量、脉宽、远场、信噪比等参数。为了判断参数测量系统的工作性能,采用均方根(RMS)误差来描述测量系统的可靠性。经过实验测试,能量测量单元的测量范围为10~1000J,标定实验数据的RMS误差为2.2%。脉宽测量单元的时间测量范围为0.5~18.0ps,时间分辨率为0.07ps,测试数据的RMS误差为3%。远场测量单元的空间测量范围为150倍衍射极限(DL),空间分辨率为0.3倍DL,测试数据的RMS误差为0.15%。信噪比测量单元的时间测量范围为30ps,时间分辨率为0.3ps,动态范围为106。基于拍瓦实验提供的测试数据表明,皮秒参数测量系统能够稳定可靠地提供以上参数的实时测试数据,实现拍瓦装置的运行状态诊断功能。 Picosecond diagnostics system is used to provide parameters of laser facility and help it to achieve designed technical aim. This system will diagnose parameters of compressed pulse for petawatt laser such as energy, pulse width, farfield, and pulse contrast. Root-mean-square (RMS) error is adopted to describe stability of diagnostics system in order to test its performance. Energy measurement range is 10-1000 J, RMS error of calibration data is 2.2 %. Pulse width measurement range is 0.5- 18.0 ps, temporal resolution is 0.07 ps for Gaussian type pulse, and data RMS error is 3 %. Farfield measurement range is 150 diffraction limitation (DL), spatial resolution is 0.3 DL, and RMS error is 0.15%. Pulse contrast measurement range is 30 ps, temporal resolution is 0.3 ps, and dynamic range is 10^6. In experiments on petawatt laser, picosecond diagnostics system provides these parameters instantaneously and successfully to analyze petawatt laser facility's performance.
出处 《中国激光》 EI CAS CSCD 北大核心 2013年第1期192-199,共8页 Chinese Journal of Lasers
基金 国家专项基金(1019631-A00)资助课题
关键词 超快光学 精密测试 自相关 单次脉冲 ultrafast optics precise diagnostics auto-correlation single shot pulse
作者简介 欧阳小平(1980-),男,博士,助理研究员,主要从事高功率激光参数测量方面的研究。E-mail:oyxp@siom.ac.cn 通信联系人。E-mail:axiulowt@hotmail.com 导师简介:朱健强(1964-),男,博士,教授,主要从事高功率激光总体技术方面的研究。E-mail:jqzhu@siom.ac.cn
  • 相关文献

参考文献21

  • 1L J. D. Zuegel, Bahk Seung-Whan, J. Bromage et al.. Novel laser and diagnostic technologies for the OMEGA EP high-energy petawatt laser[J]. Rev. Laser Engng. , 2009, 37(6):437-442.
  • 2J. Bromage, S. W. Bahk, D. Irwin et al.. A focal-spot diagnostic for on-shot characterization of high energy petawatt lasers[J]. Opt. Express, 2008, 16(21): 16561-16572.
  • 3B. E. Krusehitz, S. W. Bahk, J. Bromageet al.. Improved on- shot focal spot diagnosis on the OMEGA EP short-pulse lasersystemiCS. CLEO San Jose, CA, 2010.
  • 4夏兰,杨学东,徐光,戴亚平.光栅拼接旋转误差检测系统[J].中国激光,2011,38(3):164-168. 被引量:6
  • 5欧阳小平,张福领,唐顺兴,朱健强.短脉冲激光信噪比测量的标定方法[J].中国激光,2009,36(4):945-948. 被引量:3
  • 6J. T. Fournier, E. Snitzer. The nonlinear refractive index of glass[J].IEEE J. Quantum Electron. , 1974, QE-10(5): 473-475.
  • 7Xing Lu, Qingcao Liu, Zuoye Liu et al.. Measurement of nonlinear refractive index coefficient using emission spectrum of filament induced by gigawatt-femtosecond pulse in BK7 glass[J].Appl. Opt., 2012, 51(12): 2045-2050.
  • 8J. A. Fleck, J. R. Morris, E. S. Bliss. Small-scale self- focusing effects in a high power glass laser amplifier[J]. IEEE Journal of Quantum Electron. , 1978, QE-14(5) : 353-363.
  • 9欧阳小平,杨琳,彭永华,唐顺兴,刘崇,马金贵,钱列加,冷雨欣,朱宝强,朱健强.皮秒自相关仪的性能测试研究[J].中国激光,2012,39(4):154-157. 被引量:7
  • 10Xiaoping Ouyang, Jingui Ma, Lin Yang et al.. Accuracy of single shot autocorrelation measurements of petawatt laser pulses [J]. Appl. Opt., 2012, 51(18), 3989-3994.

二级参考文献101

共引文献43

同被引文献60

引证文献5

二级引证文献9