热释光剂量测量系统检定的质量保证及刻度因子的应用 被引量:7

Quality assurance for verification of thermoluminescence dosimetry system and application of calibration factors
原文传递
导出
摘要 目的为保证辐射受照人员个人剂量监测数据的可靠性,探讨热释光剂量测量系统检定过程影响因素,计算线性和不同能响刻度因子,确定不同类别照射群体所使用的刻度因子。方法依据标准JJG 593-2006,设计2种方案进行系统检定。方案1:已筛选退火后的探测器片放于有机玻璃平衡板中,分别予以不同剂量的137Csγ射线照射;方案2:将探测器片放于剂量计盒中,置于与方案1相同位置,予γ射线照射和窄谱过滤X射线照射。计算线性和刻度因子。结果直接照射探测器片与将探测器片放于剂量计壳中照射刻度因子差异以及不同辐射能量响应刻度因子差异均具有统计学意义。结论热释光剂量测量系统检定时应严格控制平衡条件,带剂量计盒平衡。对于放射诊断照射群体,应用低能量刻度因子;对于放射治疗照射群体,应用γ射线照射的刻度因子。2种刻度因子的使用,使监测数据更为准确。 [ Objective ] To ensure the reliability of individual dose monitoring data of radiation exposed workers, explore the inlluencing factors during verification process of thermoluminescence (TL) dosimetry system, calculate the linearity and the calibration factors of different energy respond, and determine the calibration factors that were suitable to different types of radiation exposed workers. [ Methods] According to JJG 593-2006, two schemes were designed for verification. Scheme one: the chosen TL detectors were placed directly on the phantom and irradiated by different dosage 137 Cs-γ, ray radiation. Scheme two: the chosen TL detectors were worn by boxes, then placed on the phantom and irradiated by different dosage 137Cs-γ radiation and different energy X rays. The linearity and calibration factors were calculated. [ Results] There was significant difference between the detectors placed directly on the phantom and detectors worn by boxes, and the difference between calibration factors of different energy respond was signifi- cant, too. [ Conclusion] In the course of the verification of TL dosimetry system, the balance condition should be controlled strictly, and it is important to adopt the detector boxes for balance. The low energy X-ray radiation calibration factors should be used in radiation diagnosis group, and the γ-ray radiation calibration factors is suitable to radiation therapy group. The application of two calibration factors can get more accurate monitoring data.
出处 《职业与健康》 CAS 2012年第18期2191-2193,共3页 Occupation and Health
关键词 热释光剂量测量系统 检定 质量保证 刻度因子 Thermoluminescence (TL) dosimetry system Verification Quality assurance Calibration factors
作者简介 李秀芹,女,副主任医师,主要从事放射防护监督监测工作。
  • 相关文献

参考文献2

二级参考文献8

  • 1GB18871-2002.电离辐射防护与辐射源安全标准.[S].,..
  • 2ICRU No 39. Determination of dose equivalents resulting from external radiation sources. Bethesda, MD: ICRU, 1985.
  • 3ICRU No 43. Determination of dose equivalents resulting from external radiation sources-part2. Bethesda, MD: ICRU, 1988.
  • 4ICRU No 47. Determination of dose equivalents resulting from external radiation sources. Bethesda, MD: ICRU, 1992.
  • 5Bohm J. Calibration of personal dosimeters for photon radiation with respect to the personal dose equivalent Hp(10). IAEA, Tecdoc-1126.
  • 6ICRPNo74.外照射放射防护中使用的换算系数[M].北京:原子能出版社,1998..
  • 7GB/T12162.3-2004.用于校准剂量仪和剂量率仪及确定其能量响应的X和γ参考辐射.第三部分:场所剂量仪和个人剂量计的校准及其能量响应和角响应的确定.[S].,..
  • 8GB/T12162.1-2000.用于校准剂量仪和剂量率仪及确定其能量响应的X和γ参考辐射.第一部分:辐射特性及产生方法.[S].,..

共引文献3

同被引文献46

引证文献7

二级引证文献19

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部