期刊文献+

心理护理对非接触式眼压计眼压测量值的影响

The Influence of Psychological nursing on the result of Non - contact Tonometer
在线阅读 下载PDF
导出
摘要 目的:明确心理护理对非接触式眼压计测量值的影响。方法:100例患者(200眼),分别用非接触式眼压计测量心理护理前、心理护理后眼压以及用Goldmann压平眼压计测量眼压。结果:护理前眼压为15.3±2.19mmHg,护理后眼压为14.7±2.40mmHg,减去Goldmann压平眼压计测量眼压值的差值取绝对值后分别为3.11±1.71mmHg和2.45±1.49mmHg,两者之间差异有显著性(t=2.91,P<0.05)。结论:心理护理能提高非接触式眼压计眼压测量值的准确性。 [Objective]To investigate the influence of psychology on the result of non - contact tonometer.[Method]100 patients (200 eyes) were checked by non - contact tonometer(NCT) and goldmann applantation tonometer before and after psychological nursing.[Result]The intraocular pressure(IOP)checked by NCT was 15.3±2.19mmHg before psychological nursing and the difference in the result of goldmann applantation tonometer was 3.11±1.71mmHg,The intraocular pressure were 14.7±2. 40mmHg and 2.45±1.49mmHg after psychology nursing.The difference in the result of NCT and goldmann applantation tonometer after psychology nursing was less than that before psychology nursing(t = 2.91,P<0.05).[Conclusion]Psychological nursing can improve the veracity of the result of NCT.
出处 《浙江医学教育》 2006年第4期54-55,共2页 Zhejiang Medical Education
关键词 非接触式眼压计 GOLDMANN压平眼压计 眼压测量 non-contact tonometer goldmann applantation tonometer intraocular pressure
  • 相关文献

参考文献3

二级参考文献42

  • 1Pandav SS,Sharma A,Gupta A,et al.Reliability of ProTon and Goldmann applanation tonometers in normal and postkeratoplasty eyes[J].Ophthalmology,2002,109(5):979-984.
  • 2Garzoz HJ,Chung HS,Lang Y,et al.Intraocular pressure and photorefractive keratectomy a comparison of three different tonometers[J].Cornea,2001,20(1):33-36.
  • 3Duch S,Serra A,Castanera J,et al.Tonometry after laser in situ keratomileusis treatment[J].J Glaucoma,2001,10(4):261-265.
  • 4Kaufmann C,Bachmann LM,Thiel MA.Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis[J].Invest Ophthalmol Vis Sci,2003,44(9):3790-3794.
  • 5Munoz LA,Perez AL,Rahhal MS,et al.Assessment of applanation tonometry after hyperopic laser in situ keratomileusis[J].Cornea,2002,21(2):156-160.
  • 6Copt RP,Tomas R,Mermoud A.Corneal thickness and normal tension glaucoma[J].Arch Ophthalmol,1999,117(1):14-16.
  • 7Dekker PW,Robert YCA,Kanngiesser H,et al.Principles of contact lens tonometry[J].International Ophthalmology,1998,22(2):105-111.
  • 8Shah S,Chatterjee A,Mathai M,et al.Relationship between corneal thickness and messured intraocular pressure in a general opthalmology clinic[J].Ophthalmology,1999,106(11):2154-2160.
  • 9Ventura ACS,Bohnke M,Mojon DS.Central corneal thickness measurements in patients with normal tension glaucoma,primary open angle flaucoma,pseudoexfoliation glaucoma,or ocular hypertension[J].Br J Ophthalmol,2001,85(6):792-795.
  • 10Brandt JD,Beiser JA,Kass MA,et al.Central corneal thickness in the ocular hypertension treatment study(OHTS)[J].Ophthalmology,2001,108(10):1779-1785.

共引文献50