期刊文献+

中红外无创血糖研究进展并论角质层影响 被引量:9

Progress of MIR Non-Invasive Blood Glucose Determination and Effect of Stratum Corneum
原文传递
导出
摘要 在介绍中红外无创血糖研究进展的基础上,实验探讨了组织液渗透到手指皮肤表面时,角质层对血糖测量的影响。加入角质层干扰信息前,用偏最小二乘法(PLSR)建立了模拟组织液中葡萄糖定标模型,其内部交叉验证均方差(RMSECV)为9.6 mg/dL;加入三个志愿者手指皮肤吸收的干扰信息后,所得三个模型的RMSECV分别为17.4,16.3和17.1 mg/dL。结果表明,当组织液渗透到手指皮肤表面时,不同志愿者的角质层对定标模型的影响程度大致相同;计入角质层的干扰,中红外光谱无创血糖的检测精度有望达到18 mg/dL。 The progress of non-invasive blood glucose determination based on mid-infrared spectroscopy is introduced firstly.After this,the effect of stratum corneum(SC) on blood glucose determination is studied by experiment while interstitial fluid(ISF) is collected on the skin surface.Using the spectra of simulated ISF,partial least square regression(PLSR) calibration model of glucose determination is estabished,and the value of root mean squares error of cross validation(RMSECV) is 9.6 mg/dL.While including the interference of stratum corneum from three volenteers individually,three RMSECV are obtained,and they are 17.4,16.3 and 17.1 mg/dL,respectively.The results show that the effects of SC from different volenteers are the same nearly;including the interference of SC,the accuracy of the non-invasive blood glucose determination based on attenuated total reflection mid-infrared spectroscopy can reach 18 mg/dL.
出处 《光学学报》 EI CAS CSCD 北大核心 2011年第9期58-63,共6页 Acta Optica Sinica
基金 国家自然科学基金(61078038 60938002 60878052) 中国科学院知识创新工程领域前沿(Y00232Q100) 应用光学国家重点实验室开放基金(O9Q13FQ090)资助课题
关键词 中红外光谱 衰减全反射 无创血糖检测 糖尿病 mid-infrared spectroscopy attenuated total reflection non-invasive blood glucose determination diabetics
作者简介 陈星旦(1927-),男,中国科学院院士,研究员,博士生导师。1950年毕业于湖南大学物理系。20世纪50、60年代,在物理测量方面做过多项国内急需的开创性工作。在我国第一次核爆光辐射威力测试中,创造性地提出测量方案及辐射传感、模拟、标定系统。研制的几种光冲量计,在第一次及以后历次大气层核试验中得到成功应用。20世纪80、90年代,研究真空紫外软x射线光学技术,获中国科学院科技进步一、二等奖,国家科技进步二等奖(均为第一获奖者)。近年,主要研究近红外光谱技术在农业及无创生化分析中的应用。1979年由王大珩先生指派筹办《光学学报》,临时抽调去中国科学院上海光学精密机械研究所建立学报编辑部并任第一届编委会常务副主编及编辑部主任。E—mail:chenxd@ciomp.ac.cn 通信联系人.E—mail:wdongmin@126.com
  • 相关文献

参考文献28

  • 1N. S. Oliver, C. Toumazou, A. E. C-. Cass et al.. Glucosc sensors: a review of current and emerging technology [J]. Diabetic Medicine, 2009, :16(3): 197-210.
  • 2丁海泉,卢启鹏,彭忠琦,陈星旦.近红外光谱技术用于无创生化检验研究的进展[J].光谱学与光谱分析,2010,30(8):2107-2110. 被引量:13
  • 3R. J. McNichols, G. L. Cote. Optical glucose sensing in biological fluids: an overview[J]. J. Biomed. Opt.. 2000,5(1): 5-16.
  • 4T. O. Jonathon, L. Z. Liu, V. Z. Valerie et al.. In vivo near-infrared spectroscopy of rat skin tissue with varying blood glucose levels [J]. Anal. Chem., 2006, 78(1): 215-223.
  • 5P. Roychoudhurya, L. M. Harveya, B. McNeil. At -line monitoring of ammonium, glucose, methyloleate and biomass in a complex antibiotic fermentation process using attenuated total reflectance-mid infrared (ATR-MIR) spectroscopy[J]. Analytica Chimica Acta., 2006, 561(1 -2):218-224.
  • 6N. Kaiser. Laser absorption spectroscopy with an ATR prism [J]. IEEE TransBiomed Engng. , 1979, 26(10): 597-600.
  • 7R. Vonach, J. Buschmann, R. Falkowski et al.. Application of mid-infrared transmission spectrometry to the direct determination of glucose in whole blood[J]. Appl. Spectrosc. , 1998, 52(6):820-822.
  • 8G. Budinova, J. Salva, K. Volka. Application of molecular spectroscopy in the mid infrared region to the determination of glucose and cholesterol in whole blood and in blood serum [J]. Appl. Spectrosc., 1997, 51(5): 631-635.
  • 9K. J. Ward, D. M. Haaland, M. R. Robinson et al.. Postprandial blood glucose determination by quantitative mid-infrared spectroscopy[J]. Appl. Spectrosc., 1992, 46(6): 959-965.
  • 10W. B. Martin, S. Mirov, R. Venugopalan. Using two discrete frequencies within the middle infrared to quantitatively determine glucose in serum [J]. J. Biomed. Opt. , 2002, 7(4) : 613-617.

二级参考文献21