Determination of Amantadine Residue in Honey by Solid-phase Extraction and High-performance Liquid Chromatography with Pre-column Derivatization and Fluorometric Detection 被引量:15

Determination of Amantadine Residue in Honey by Solid-phase Extraction and High-performance Liquid Chromatography with Pre-column Derivatization and Fluorometric Detection
原文传递
导出
摘要 Amantadine (AMA) is an anti-viral drug used in apiculture to protect honeybee against the sacbrood virus (Morator aetatulae). This study described a reliable high-performance liquid chromatographic (HPLC) method for analyzing AMA in honey using a solid-phase extraction (SPE) cartridge (Plexa PCX) for purification, 4-fluoro-7- nitro-2,1,3-benzoxadiazole (NBD-F) as a pre-column derivatization agent, and fluorometric detection (λex =470 nm, λem=530 nm). The chromatographic separation was performed on an XDB C18 column (150×4.6 mm i.d.) using 0.1% trifluoroacetic acid/acetonitrile (35 ; 65, V/V) as the mobile phase at a flow rate of 1.0 mLomin 1 with a run time of 20 min. Under these optimal conditions, a linear relationship was observed in the range of 0.025--1.0μg·mL-1 with a good correlation coefficient (0.998) and low limit of detection (0.0080 μg·g-1), the recoveries were all above 90%, and the intra-day and inter-day precision (RSD) ranged from 3.4%--5.1%. Amantadine (AMA) is an anti-viral drug used in apiculture to protect honeybee against the sacbrood virus (Morator aetatulae). This study described a reliable high-performance liquid chromatographic (HPLC) method for analyzing AMA in honey using a solid-phase extraction (SPE) cartridge (Plexa PCX) for purification, 4-fluoro-7- nitro-2,1,3-benzoxadiazole (NBD-F) as a pre-column derivatization agent, and fluorometric detection (λex =470 nm, λem=530 nm). The chromatographic separation was performed on an XDB C18 column (150×4.6 mm i.d.) using 0.1% trifluoroacetic acid/acetonitrile (35 ; 65, V/V) as the mobile phase at a flow rate of 1.0 mLomin 1 with a run time of 20 min. Under these optimal conditions, a linear relationship was observed in the range of 0.025--1.0μg·mL-1 with a good correlation coefficient (0.998) and low limit of detection (0.0080 μg·g-1), the recoveries were all above 90%, and the intra-day and inter-day precision (RSD) ranged from 3.4%--5.1%.
出处 《Chinese Journal of Chemistry》 SCIE CAS CSCD 2011年第8期1764-1768,共5页 中国化学(英文版)
关键词 AMANTADINE HONEY solid-phase extraction high-performance liquid chromatographic (HPLC) fluorometric detection amantadine, honey, solid-phase extraction, high-performance liquid chromatographic (HPLC),fluorometric detection
作者简介 E-mail: apiswu@ 126.com; Tel.: 0086-010-82595649; Fax: 0086-010-82595649
  • 相关文献

参考文献23

  • 1Arndt, T.; Guessregen, B.; Hohl, A.; Reis, J. Clin. Chim. Acta 2005, 359, 125.
  • 2Schwab, R. S.; England, A. C. Jr.; Poskanzer, D. C.; Young, R. R. J. Am. Med. Assoc. 1969, 208, 1168.
  • 3Smorodintsev, A. A.; Zlydnikov, D. M.; Kiseleva, A. M.; Romanov, J. A.; Kazantsev, A. P.; Rumovsky, V. I. J. Am. Med. Assoc. 1970, 213, 1448.
  • 4Bright, R. A.; Shay, D. K.; Shu, B.; Cox, N. J.; Klimov, A. I. J. Am. Med. Assoc. 2006, 295, 891.
  • 5Cheung, C. L.; Rayner, J. M.; Smith, G. J.; Wang, P.; Nai- pospos, T. S.; Zhang, J.; Yuen, K. Y.; Webster, R. G.; Peiris, J. S.; Guan, Y.; Chen, H. J. Infect. Dis. 2006, 193, 1626.
  • 6He, G.; Qiao, J.; Dong, C.; He, C.; Zhao, L.; Tian, Y. Antiviral Res. 2008, 77, 72.
  • 7FDA Veterinarian Newsletter, Vol. XXI, U.S. Department of Health and Human Services Public Health Service, Food and Drug Administration, 2006, p. 6.
  • 8Yang, Z. Z. J. Bee 1990, 20(7), 3.
  • 9Grabensteiner, E.; Ritter, W.; Carter, M. J.; Davison, S.; Pechhacker, H.; Kolodziejek, J.; Boecking, O.; Derakhshifar, I.; Moosbeckhofer, R.; Licek, E.; Nowotny, N. Clin. Diagn. Lab. Immunol. 2001, 8(1), 93.
  • 10Ying, S. T.; Sun, C. X.; Dong, G. Q.; Zhang, Q. H. Apicuture of China 2008, 59(7), 12.