摘要
通过观察研究脉诊仪在中医诊断实验教学中的使用情况,作者发现中医脉诊仪在使用过程中存在几个应该注意的问题:(1)寸关尺三部指示灯不亮,电子芯片不稳定;(2)诊脉部位不明确;(3)仿真手指容易断裂;(4)设备运行时噪音太大;(5)机身后侧闪烁灯暴露;(6)配套的计算机软件不稳定。改进对策:(1)加强寸关尺三关芯片与仿生手臂的黏合程度,或将芯片与三关部位嵌合为一整体;(2)将寸关尺三关、桡动脉脉道与仿真手臂三者嵌合固结为一体;(3)在手指内部置入钢性介质或者其他韧性的类似骨架介质,更换成其他耐用耐磨的材料;(4)在多方面采取降噪措施;(5)设备生产时将闪烁指示灯密封于机器内部,或者有专门的灯罩将其遮盖;(6)应该强化技术进一步提高计算机软件的稳定性和可行性。
Observation usage of pulse meter in traditional Chinese medicine diagnosis experimental teaching,we found some problems that should be noted:(1)LED does not light,electron Chip unstable;(2)diagnostic parts is not clear;(3)simulation fingers easily are broken;(4)the noise is large;(5)Blinking light is easily exposure;(6)Supporting computer software is unstable.Countermeasures:(1)the relevant parts of the chip with the three chimeric as a whole;(2)Three in one(3)Placed inside the finger or other medium toughness of steel similar skeleton medium replaced with other durable wear-resistant materials.(4)noise reduction measures should be taken in many ways;(5)The equipment will be sealed inside the machine flashing lights,or have a special shade to cover;(6)Improving the stability and feasibility of computer software.
出处
《辽宁中医药大学学报》
CAS
2011年第6期271-272,共2页
Journal of Liaoning University of Traditional Chinese Medicine
关键词
脉诊仪
实验教学
存在问题
对策
pulse meter
experimental teaching
problems
countermeasures
作者简介
张丽娜(1982-),女,辽宁沈阳人,助理实验师,硕士,研究方向:临床技能实验准备及研究工作。