期刊文献+

ICP-AES测定铜合金中的磷 被引量:10

Determination of Phosphorus in Copper Alloy by ICP-AES
在线阅读 下载PDF
导出
摘要 采用ICP-AES测定铜合金中的磷元素。通过实验探讨了铜合金中基体元素,共存元素对磷元素分析谱线的光谱干扰情况,确定了合适的分析谱线。对一系列不同牌号不同含量的铜合金中的磷进行测定,通过实验探讨了铜合金中基体元素铜对磷元素测定的干扰情况。方法的检出限为0.00084%,Z-15铍青铜和铅黄铜测定结果的相对标准偏差分别为2.06%和2.89%。加标回收率为85.3%—98.2%。 The phosphorus in copper alloy was determined by ICP-AES.The spectral interference of matrix and coexist elements were studied,and the analytical line was selected.Phosphorus in copper alloy which had a series of various brands and different content was determined.The interference of copper substrate on the determination of phosphorus was studied.The detection limit of the method was 0.00084%.In application of the method to the determination of Z-15 beryllium-copper and lead brass,the relative standard deviations were 2.06% and 2.89%,and the recovery of 85.3%—98.2%.
出处 《光谱实验室》 CAS CSCD 北大核心 2010年第3期1151-1155,共5页 Chinese Journal of Spectroscopy Laboratory
关键词 电感耦合等离子体-原子发射光谱法 铜合金 光谱干扰 ICP-AES Copper Alloy Phosphorus Spectral Interference
作者简介 联系人,电话:(010)62496661;手机:(0)15011519152;E—mail:erhong_fude@163.com 付二红(1982-),女,河北省保定市人,助理工程师,硕士,主要从事化学分析工作。
  • 相关文献

参考文献3

共引文献9

同被引文献460

引证文献10

二级引证文献45

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部