利用源皮距验证头部肿瘤放疗摆位误差可行性的研究

Study of Feasibility of Verifying Setup Errors for Patients with Head Tumor by Observing Patients’SSD
在线阅读 下载PDF
导出
摘要 目的验证源皮距法保证头部肿瘤患者精确摆位的可行性。方法使用Varian Acuity8.6模拟机对21例头部肿瘤病人拍片,并从治疗计划系统中得到相应位置的标准射野数字重建(DRR)图像,应用Acuity的二维匹配功能,得出摆位误差;观察其与射野角度为0°.45°.315°时源皮距误差的对应关系。结果21例患者左右、头脚、前后方向系统误差分别为(1.1±1.6)mm、(0.7±1.2)mm、(0.9±1.5)mm,随机误差分别为(1.51±3.1)mm、(1.05±3.3)mm、(1.60±2.3)mm。各角度源皮距系统误差为(1.25±1.3)mm、(1.04±1.3)mm、(1.10±2.3)mm,随机误差为(2.03±1.7)mm、(2.81±2.3)mm、(2.33±3.0)mm。结论源皮距法简易可行,可以作为其他位置验证手段的辅助。 Objective To verify the feasibility of setup error verification by observing patents' source-skin distance(SSD) for patients of head tumor. Methods Films for 21 patients with head tumor were recorded using simulator (Varian Acuity8.6), and comparison with reference digitally reconstructed radiograph (DRR) from Treatment plan system (TPS). The deviation of setup for 21 patients in the left-right, anterior-posterior and superior-inferior directions were measured by using 2D match, and SSD error was recorded when gantry angle was 0°, 45°, 315°. Then setup error and corresponding SSD error were analyzed. Results The systematic errors and random errors of 21 patients in the left-right, anterior-posterior and superior-inferior directions were (1.1 ± 1.6)mm, (0.7± 1.2)mm, (0.9 ± 1.5)mm and ( 1.51 ± 3.1 ) mm, (1.05± 3.3) mm, (1.60 ±2.3)mm .The systematic SSD errors and random SSD errors were (1.25 ± 1.3)mm (1.04 ± 1.3)mm, (1.10±2.3)mm, and (2.03± 1.7)mm, (2.81 ±2.3)mm, (2.33±3.0)mm for gantry angle was 0°, 45° 315°,respectively. Conclusions It is simple and feasible for setup error verification by observing patients' SSD and can be auxiliary to other verification means.
出处 《中国医疗器械杂志》 CAS 2010年第1期60-61,共2页 Chinese Journal of Medical Instrumentation
关键词 头部肿瘤 放射治疗 摆位误差 源皮距 head tumor, radiotherapy, setup error, source-skin distance
作者简介 崔恩萍,主管技师,E-mail:tanghong92500@yahoo.com.cn
  • 相关文献

参考文献3

  • 1Ranald I, Paul A, John P, et al. Patient positioning using detailed three-dimensional surface data for patients undergoing conformal radiation therapy for carcinoma of the prostate: A feasibility study[J]. Int. J. Radiation Oncology Biol. Phys.2001,49: 225-230.
  • 2Gilbeau L, Octave-Prignot M, Loncol T, et al. Comparison of setup accuracy of three different thermoplastic masks for the treatment of brain and head and neck tumors[J]. Radiotherapy and Oncology. 2001,58: 155-162.
  • 3肖锋,孙朝阳,胡明民,陆宙,刘晓莉,陆军,石梅.三维适形及调强放疗摆位误差分析[J].中国医学物理学杂志,2008,25(3):641-642. 被引量:21

二级参考文献9

  • 1Prescribing recording and reporting photon beam therapy [R]. ICRU. Report no.50. 1993.
  • 2Prescribing recording and reporting photon beam therapy (Supplement to ICRU report 50) [R]. ICRU. Report no. 62. 1999.
  • 3Van Herk M, Remeijer P, Lebesque Jr. Inclusion of geometric uncertainties in treatment plan evaluation [J]. Int J Radiat Oncol Biol Phys, 2002,52(5):1407-1422.
  • 4Mckenzie A, Van Herk M, Mijnheer B. Margins for geometric uncertainty around organs at risk in radiotherapy [J]. Radiotherapy and Oncology, 2002,62(3):299-307.
  • 5Hurkmans CW, Remeijer P, Lebesque JV, et al. Set-up verification using portal images review of current clinical practice [J]. Radiotherapy and Oncology, 2001,58(2):105-120.
  • 6Stroom JC, de Boer HC, Huizenga H, et al. Inclusion of geometrical uncertainties in radiotherapy treatment planning by means of coverage probability [J]. Int J Radiat Oncol Biol Phys. 1999,43(4):905-919.
  • 7Joseph Hanley, Marc M. Debois, Dennis Mah, et al. Deep inspiration breath-hold technique for lung tumors: the potential value of target immobilization and reduced lung density in dose escalation [J]. Int J Radiat Oncol Biol Phys. 1999,45(3):603-611.
  • 8Kenneth E. Rosenzweig, Joseph Hanley, Dennis Mah, et al. The deep inspiration breath-hold technique in the treatment of inoperable non-small-cell lung cancer [J]. Int J Radiat Oncol Biol Phys, 2000, 48(1): 1-87.
  • 9Joshua J. Haslam, Anthony E. Lujan, Arno J. Mundt, et al. Setup errors in patients treated with intensitymodulated whole pelvic radiation therapy for gynecological malignancies [J]. Medical Dosimetry. 2005, 30(1):36-42.

共引文献20

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部