期刊文献+

高效率快裂变室设计方案 被引量:1

A Design About Fast And High Detection Efficiency Fission Chamber
在线阅读 下载PDF
导出
摘要 介绍了一支用于瞬发中子衰减常数测量的高效率快裂变室的设计方案。裂变室采用21层平板型电极结构设计,镀层235U质量350mg,对快中子的探测效率达到1‰。采用极板窄间距设计和5路信号输出方式实现裂变室的快时间响应,其通过142型电荷灵敏前置放大器输出的信号上升时间约16ns。 A design is given about fast and high detection efficiency fission chamber for prompt neutron decay constant measurements. The chamber has 21 foils containing 350mg of 235U whose fast neutron detection efficiency is up to 1 . The fast time relation of the chamber is performed by choosing the suitably small distance between the foils and dividing the foils into 5 sections whose signals are lead out separately to 142 preamplifiers. The rise time of the signals from preamplifier is about 16ns.
出处 《核电子学与探测技术》 CAS CSCD 北大核心 2009年第2期251-254,共4页 Nuclear Electronics & Detection Technology
关键词 裂变室 上升时间 堆积 fission chamber, rise time, pileup
作者简介 周浩军(1979-),男,河北唐山人,研究实习员,从事反应堆核参数测量工作。
  • 相关文献

参考文献5

  • 1ORTEC Model 142A, 142B, and 142C Preamplifiers Operating and Service Manual
  • 2李建胜,张翼,金宇,李润良.^(252)Cf快裂变室研制[J].核电子学与探测技术,2001,21(4):264-267. 被引量:26
  • 3H. Rosler, J. K. Millard, N. W. Hill. Nuclear Instruments and Methods, 1972,99.477-486.
  • 4A.A. Bogdzel. Nuclear Instruments and Methods, 1982,200 : 407-410.
  • 5A. A. Bogdzel, N. A. Gundorin. Nuclear Instruments and Methods in Physics Research A, 1994, 343 : 545-549.