编码激励超声成像系统中二进制最优编码序列的研究 被引量:11

Study of Optimal Binary Sequences in Ultrasound Imaging System Using Coded Excitation
在线阅读 下载PDF
导出
摘要 本研究对医学超声成像中可能应用的二进制最优编码序列进行了深入研究。提出了应用于医学超声的二进制最优及近似最优编码序列及其判断标准。通过穷举搜索给出了长度在15—33位的最优及近似最优编码序列,并选取长度为25的近似最优序列进行实验研究,与长度为13位的Barker码进行对比实验,证实了本研究所提出的二进制最优编码序列的设计方法是有效的。与Barker码相比,使用增加长度的最优序列进行编码激励可以获得更高的信噪比增益,同时保持与Barker码相近的距离旁瓣水平。 The optimal binary sequences used in medical ultrasound imaging were studied in this paper. The concept and the criteria of optimal binary sequences used in medical ultrasound imaging were proposed. Optimal sequences of length 15-33 were obtained through exhaustive search. An optimal sequence of length 25 was chosen to take comparative experiments with Barker code of length 13, and the effectiveness of optimal sequences was proved. Compared with Barker codes, the longer optimal binary sequences provided higher SNR gain, while similar range sidelobe level was remained.
作者 刘凯 高上凯
出处 《中国生物医学工程学报》 CAS CSCD 北大核心 2007年第1期42-47,共6页 Chinese Journal of Biomedical Engineering
关键词 编码激励 二进制编码序列 医学超声 coded excitation binary code sequence medical ultrasound
作者简介 通讯作者。E-mail:gsk-dea@tsinghua.edu.cn
  • 相关文献

参考文献10

  • 1O' Donnen M. Coded Excitation System for Improving the Penetration of Real-Time Phased-Array Imaging Systems [J] . IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control,1992, 39(3): 341-351.
  • 2Welch LR, Fox MD. Practical spread spectrum pulse compression for ultrasonic tissue imaging[J].IEEE Transactions on Ultrasonics,Ferroelectrics, and Frequency Control, 1998, 45:349- 355.
  • 3Ruprecht J, Rupf M. On the search for good aperiodic binary invertible sequences [J]. IEEE Trans. Inform. Theory, 1996, 42:1604- 1612.
  • 4Cohen MN, Fox MR, Baden JM. Minimum peak sidelobe pulse compression codes [A]. Proc. LEEE Int. Radar Conf [C]. 1990.633 - 639.
  • 5Deng XM, Fan PZ. New binary sequences with good aperiodic autocorrelations obtained by evolutionary algorithm [ J ]. IEEE Communications Letters, 1999, 3: 288-290.
  • 6Chiao RY, Hao XH. Coded excitation for diagnostic ultrasound: A system developer's perspective [ J ]. IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2005, 52(2) :160- 170.
  • 7Liu K, Gao SK. Searching binary sequences for coded excitation in medical ultrasound [ A]. In: 27th Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings[C]. Shanghai: IEEE, 2005. 1858- 1860.
  • 8Hayes MH. Statistical Digital Signal Processing and Modeling [M].New York: Wiley, 1996. 166-174.
  • 9Ebbini ES, Li PC, Shen J. A new SVD-based optimal inverse filter design for ultrasonic applications [A]. In: Proceedings of 1993 Ultrasonics symposium [C]. Baltimore, MD: IEEE, 1993. 1187-1190.
  • 10Liu K, Peng QY, Wang XQ, et al. Vitreous image enhancement in an ophthalmic sonographic imaging system using Golay coded excitation[J]. Journal of Ultrasound in Medicine, 2005, 24: 1251 - 1259.