X光高分辨探测用CsI(Tl)晶体的蒙特卡罗模拟研究 被引量:5

Study on CsI(Tl) Scintillating Crystal for X-Ray High-Resolution Detection with Monte Carlo Method
原文传递
导出
摘要 提出了一种基于CsI(Tl)闪烁晶体和面阵CCD器件、采用光纤和光纤面板进行光耦合及传输、以扇形束线阵扫描方式实现对X光高分辨探测的方案。CsI(Tl)晶体的尺寸大小将直接影响到晶体的发光效率及X光的高分辨探测,据此开展了蒙特卡罗模拟研究。模拟研究了X射线能量、X射线源到探测晶体的距离(源距)、CsI(Tl)晶体的厚度与X射线能量分布、全能峰效率与CsI(Tl)闪烁晶体转换效率之间的关系。结果表明,当X射线能量为120~450keV,CsI(Tl)晶体尺寸厚度为0~1.5cm变化时,全能峰效率的变化范围为31.34~96.74%,CsI(Tl)闪烁晶体的转换效率的变化范围为12.8~97.43%。可见,X射线的能量及CsI(Tl)闪烁晶体尺寸的厚度,是决定X光高分辨探测的重要参量,这对优化X光高分辨探测用CsI(Tl)晶体的尺寸设计具有一定的参考价值。 A X-ray detector, based on the CsI(Tl) scintillating crystal and area-array CCD, using fiber and fiber optic plate to couple and transmit light with fan-beam linear array scanning, has been put forth. The luminescent efficiency and X-ray high-resolution detection are directly influenced by the dimension of CsI(Tl) scintillating crystal, and Monte Carlo method is adopted to simulate the influences, including the relations between the energy of X-ray, the distance between X-ray source and the scintillating crystal, and the thickness of CsI(Tl) crystal and the X-ray energy distribution, and then the whole energy peak efficiency and the transition efficiency. The results show that while the X-ray energy is of 120-450 keV, and the thickness of CsI(Tl) crystal is 0.3-1.5 cm, the whole energy peak is within the range of 31.34 % -96.74 %, and the transition efficiency is within the range of 12.8 % - 97.43 %. So, the X-ray energy and the dimension of CsI (Tl) scintillating crystal are the most important parameters determining the X-ray high resolution detection. It has certain referenced values to optimize the dimension design of CsI(Tl) scintillating crystal for X-ray high resolution detection.
出处 《光学学报》 EI CAS CSCD 北大核心 2006年第9期1429-1434,共6页 Acta Optica Sinica
基金 国家自然科学基金(60172074)资助课题
关键词 X射线光学 工业计算机X射线层析术无损检测技术 X射线高分辨探测 CsI(Tl)闪烁晶体 蒙特卡罗方法 X-ray optics industrial X-ray computerized tomography non-destructive testing X-ray high-resolutiondetection, CsI(Tl) scintillating crystal, Monte Carlo method
作者简介 魏彪(1963-),男,浙江桐乡人,重庆大学教授,博士后,主要从事光电检测与成像技术方面的研究。E-mail:weibiao1963@163.com。
  • 相关文献

参考文献18

  • 1美国无损检测手册(射线卷)[M].美国无损检测学会编,中国无损检测学会翻译,上海:世界图书出版社,2000.618~653
  • 2K. A. Spencer. Computer tomography- An overview[J].Photographic Science, 1989, 37(2): 84-89
  • 3W. B. Gilboy. X-and γ-ray tomography in NDE applications[J].Nuclear Instruments and Methods, 1984, A221: 193-200
  • 4T. Luthi, A. Flisch, P. Wyss. Industrial computed X-ray tomography[J]. NDT International, 1998, 40(3): 196-197
  • 5M. S. Rapaport, A. Gayer, E. Iszak et al.. A dual-mode industrial CT[J]. Nuclear Instruments and Methods, 1995,A352:652-658
  • 6P. Ruegsegger, B. Koller, R. Muller. A microtomographyic system for the non destructive evaluation of bone architecture[J]. Calcified Tissue Int. , 1996, 58(1): 24-29
  • 7Roy Clarke. CCD X-ray detectors: opportunities and challenges[J]. Nuclear Instruments and Methods, 1994, A347:529-533
  • 8A. Owens, Kieran J. McCarehy. Energy deposition in X-ray CCDs and charged particle discrimination [J]. Nuclear Instruments and Methods, 1995, A366:148-154
  • 9Tetsuhiko Takhoshi, Haruo Iton, Toshikazu Shimado et al..Design of integrated radiation detectors with a-Si photodiodes on ceramic scintillators for use in X-ray CT[J]. IEEE Transactions on Nuclear Science, 1990, 37(3): 1478-1503
  • 10K. Machin, S. Webb. Cone-beam X-ray microtomography of small samples[J]. Phys. Med. Biol., 1994, 39(7): 1639-1657

二级参考文献26

  • 1谢红兰,陈建文,高鸿奕,熊诗圣,李儒新,徐至展.X射线荧光全息术中入射能量对原子像的影响[J].光学学报,2004,24(7):1004-1008. 被引量:3
  • 2Yao Yuqi,J Opt Soc Am A,1997年,14卷,1期,325页
  • 3Das B B,Rep Prog Phys,1997年,60卷,227页
  • 4Su Chang,博士学位论文,1996年,45页
  • 5Wang Lihong,Monte Carlo Modeling of Light Transport in Multi-Layered Tissues in Standard C(Report),1992年
  • 6Cheong Waifung,IEEE J Quantum Electron,1990年,26卷,12期,2166页
  • 7Hidetaka Arimura, Takuji Date, Kaoru Morikawa et al.. Effect of scattered X-rays on image signal of radiograph[C]. Proc. 80^th Conference of Japan Society of Medical Physics, 2000. 33-41.
  • 8Dinko. E. Gonzalez Trotter, J. Eric. Tkaczyk, John Kaufhold et al.. Thickness-dependent scatter correction algorithm for digital mammography[C]. Proc. SPIE. 2002, 4682:469-478.
  • 9J. A. Seibert, J. M. Boone. X ray scatter removal by deconvolution[J]. Med. Phys., 1988, 15(4): 567-575.
  • 10John M. Boone, J. Anthory Seibert. An analytical model of the scattered radiation distribution in diagnostic radiology[J]. Med.Pkys., 1988, 15(5): 721-725.

共引文献35

同被引文献50

  • 1江孝国,王伟,王婉丽.X光作用下CsI:Tl晶体的转换效率研究[J].强激光与粒子束,2005,17(2):203-205. 被引量:13
  • 2邵军明,路宏年,蔡慧.X射线成像的一种点扩展函数模型[J].光学学报,2005,25(8):1148-1152. 被引量:9
  • 3陈志强,李亮,冯建春.高能射线工业CT最新进展[J].CT理论与应用研究(中英文),2005,14(4):1-4. 被引量:19
  • 4魏彪,周密,靳冬欢,冯鹏,米德伶.狭长CsI(Tl)闪烁体发光效率的研究[J].光子学报,2006,35(1):41-46. 被引量:3
  • 5周日峰,王珏,韩忠.X射线工业CT散射修正[J].原子能科学技术,2006,40(4):450-452. 被引量:6
  • 6Miyai H,Satou K,Kitaguchi H,et al.Response of silicon detector for high energy X-ray computed tomography[J].IEEE Trans On Nucl Sci,1994,41(4):999-1003.
  • 7Nuzzo S,Peyrin F,Martin E,et al.Assessment of Bone Mineral Content from 3-D Synchrotron Radiation Microtomography Images[J].IEEE Trans On Nucl Sci,2001,48(3):859-863.
  • 8T. J. Davis, D. Gao, T. E. Gureyev et al.. Phase-contrast imaging of weakly absorbing materials using hard X-rays[J]. Nature, 1995, 373(6515): 595-598
  • 9Atsushi Momose, Tohoru Takeda, Yyji Itai et al.. Phase-contrast X-ray computed tomography for observing biological soft tissues[J]. Nature Med., 1996, 2: 473-475
  • 10Franz Pfeiffer, Timm Weitkamp, Oliver Bunk et al.. Phase retrieval and differential phase-contrast imaging with low-brilliance X-ray sources[J]. Nature Physics, 2006, 2: 258-261

引证文献5

二级引证文献14

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部