期刊文献+

血管内超声图像序列的自动边缘检测 被引量:2

Automated Contour Detection forIntravascular Ultrasound Image Sequences
在线阅读 下载PDF
导出
摘要 血管内超声成像可以显示血管内腔、管壁清楚的实时截面图像,从而使定量测量血管参数成为可能。传统手动追踪内腔和管壁边界的方法既费时又有较大的主观误差,本文基于快速主动轮廓算法模型提出一种适合血管内超声图像的自动边缘检测方法。实验结果发现,自动检测的内腔、血管壁面积与手动追踪非常吻合,具有较高的相关系数(0.94和0.95)和较小的系统误差(-0.32和0.56),可作为医生可靠而准确的诊断工具。 Intravascular ultrasound can provide clear real-time cross-sectional images including lumen and plaque. Quantitative measurement can be performed by these images. Traditionally, the luminal border and medial-adventitial border are manually traced. This process is time-consuming and the subjective difference is larger. A new automated contour detection method is used for intravascular ultrasound images based on a fast active contour model. Experimental results show that luminal and vessel area measurements after automated detection are in agreement with manual trace with high correlation coefficients (0 94 and 0 95) and small system difference (-0 32 and 0 56). Thus it is used as a reliable and accurately diagnostic tool.
出处 《南京航空航天大学学报》 EI CAS CSCD 北大核心 2005年第3期313-318,共6页 Journal of Nanjing University of Aeronautics & Astronautics
关键词 边缘检测 图像处理 可变形模板 血管内超声 内腔边界 contour detection image processing medial-adventitial border intravascular ultrasound luminal border
  • 相关文献

参考文献14

  • 1Nissen S E, Gurley J C, Grines C L, et al. Intravascular ultrasound assessment of lumen size and wall morphology in normal subjects and patients with coronary artery disease[J]. Circulation, 1991,84:1087~1099.
  • 2Mallery J A, Tobis J M, Griffith J, et al. Assessment of normal and atherosclerotic arterial wall thickness with an intravascular ultrasound imaging catheter[J]. Amer Heart, 1990, 119:1392~1400.
  • 3Meier D S, Cothren R M, Vinece D G, et al. Automated morphometry of coronary arteries with digital image analysis of intravascular ultrasound[J]. Amer Heart, 1997,133:681~690.
  • 4Zhang X, McKay C R, Sonka M. Tissue characterization in intravascular ultrasound images[J]. IEEE Trans Med Imaging, 1998,17:889~899.
  • 5Haas C, Ermert H, Holt S, et al. Segmentation of 3D intravascular ultrasonic images based on a random field model[J]. Ultrasound Med Biol, 2000,26:279~306.
  • 6von Birgelen C, Mario C, Li W, et al. Morphometric analysis in three-dimensional intracoronary ultrasound: an in vitro and vivo study performed with a novel system for the contour detection of lumen and plaque[J]. Am Heat, 1996, 132:516~527.
  • 7Kovalski G, Beyar R, Shofti R, et al. Three-dimensional automatic quantitative analysis of intravascular ultrasound images[J]. Ultrasound Med Biol, 2000,26:527~537.
  • 8Kass M, Witkin A, Terzopoulos D. Snakes: active contour models[A]. Proc 1st Int Conf Computer Vision[C]. Lodon: IEEE, 1987.259~268.
  • 9Amini A A, Tehrani S, Weymouth T E. Using dynamic programming for minimizing the energy of active contours in the presence of hard constraints[A]. Proc 2nd Int Conf Computer Vision[C]. Tarpon Springs: IEEE, 1998.95~99.
  • 10Lbregt S, Viergever M A. Discrete dynamic contour model[J]. IEEE Trans Med Imag, 1995,14:12~24.

二级参考文献2

共引文献41

同被引文献27

  • 1Nissen S E,Gurley J C,Grines C L,et al.Intravascular ultrasound assessment of lumen size and wall morphology in normal subjects and patients with coronary artery disease[J].Circulation,1991,84 (3):1087 ~ 1099.
  • 2Mallery J A,Tobis J M,Griffith J,et al.Assessment of normal and atherosclerotic arterial wall thickness with an intravascular ultrasound imaging catheter[J].American Heart Journal,1990,119 (6):1392 ~ 1400.
  • 3Dejesus S T,Rosenfield K,Gal D,et al.Three-dimensional reconstruction of vascular lumen from images recorded during percutaneous 2-D intravascular ultrasound[J].Clinical Research,1998,37:838 ~838.
  • 4Jacob G,Noble J A,Blake A.Robust contour tracking in echocardiaographic sequences[A].In:Proceedings of IEEE 6th International Conference on Computer Vision[C],Bombay,India,1998:408 ~413.
  • 5Lengyel J,Greenberg D P,Popp R.Time-dependent threedimensional intravascular ultrasound[A].In:Proceedings of ACM SIGGRAPH[C],New York,1995:457 ~ 464.
  • 6von Birgelen C,de Vrey E A,Mintz G S,et al.Ecg-gated threedimensional intravascular ultrasound-feasibility and reproducibility of the automated analysis of coronary lumen and atherosclerotic plaque dimensions in humans[J].Circulation,1997,96 (9):2944 ~ 2952.
  • 7Meier D S,Cothren R M,Vinece D G,et al.Automated morphometry of coronary arteries with digital image analysis of intravascular ultrasound[J].American Heart,1997,133 (6):681~690.
  • 8Zhang X,McKay C R,Sonka M.Tissue characterization in intravascular ultrasound images[J].IEEE Transactions on Med Imaging,1998,17(6):889 ~ 899.
  • 9Haas C,Ermert H,Holt S,et al.Segmentation of 3D intravascular ultrasonic images based on a random field model[J].Ultrasound Med Biol,2000,26:279 ~ 306.
  • 10von Birgelen C,Mario C,Li W,et al.Morphometric analysis in three-dimensional intracoronary ultrasound:an in vitro and vivo study performed with a novel system for the contour detection of lumen and plaque[J].American Heart,1996,132(3):516 ~527.