期刊文献+

高纯氧化钬的电感耦合等离子体原子发射光谱分析及光谱干扰的校正 被引量:3

Determination of Trace Rare Earth Element Impurities in High Purity Holmium oxide by Inductively Coupled Plasma-Atomic Emission Spectrometry and Correction of Spectral Interferences
在线阅读 下载PDF
导出
摘要 本文提出了以PGS-2型平面光栅摄谱仪与Plasma Therm ICP-5000D射频发生器联用,乙醇溶液预去溶进样方法,直接同时测定高纯氧化钬中5个稀土杂质元素的方法,并讨论了基体浓度对分析方法检出限的影响和光谱干扰及其校正。当样品溶液中稀土总浓度为5mg/ml时,测定下限分别为铽0.003%,镝、铒和铥0.002%,钇0.0003%。其相对标准偏差为2.8%~7.4%。 The direct simultaneous determination of five trace rare earth impurities in high purity holmium oxide was described. The sample solution containing ethanol was introduced into the plasma with desolva-tion. The effect of matrix concentration on the detection limit, and the calculation and correction of spectral interferences in ICP-AES have been investigated. For the sample solution containing 5 mg/ml of Ho2O3. the determination limit was 0. 003% for Tb4O7, 0. 002% for Dy2O3, Er2O3 and Tm2O3 and 0. 0003% for Y2O3, respectively, the relative standard deviation was in the range of 2.8%-7. 4%.
出处 《分析化学》 SCIE EI CAS CSCD 北大核心 1993年第8期918-920,共3页 Chinese Journal of Analytical Chemistry
关键词 氧化钬 光谱干扰 稀土杂质元素 ICP Holmium, Terbium, Dysprosium, Erbium, Thulium, Yttrium, Inductively coupled plasma, Spectral interference.
  • 相关文献

参考文献5

二级参考文献5

  • 1杨金夫,分析化学,1988年,16卷,2期,117页
  • 2杨金夫,分析化学,1988年,16卷,3期,198页
  • 3綦文娣,稀有金属,1988年,7卷,1期,59页
  • 4张赣南,分析化学,1986年,14卷,9期,703页
  • 5黄本立,分析化学,1980年,8卷,5期,416页

共引文献3

同被引文献111

引证文献3

二级引证文献9