流动注射测定饮用水中挥发性酚类化合物方法的研究 被引量:7

Detection of volatile phenolics in drinking water by flow injection
在线阅读 下载PDF
导出
摘要 〔目的〕为获得一种方便、准确和灵敏测定水中挥发性酚类化合物的检测方法。〔方法〕应用流动注射技术在线测定饮用水中挥发性酚类化合物 ,水样在线蒸馏后不经氯仿萃取直接进行定量分析。〔结果〕方法的最低检测浓度为 1.5 μg/L :线性范围 1.5 μg L~ 5 0 μg/L呈直线 ,其相关系数r≥ 0 .999;相对标准偏差为 2 .3 %~ 7.4% ;加标回收率为 88.9%~ 10 9%。〔结论〕本方法操作简便、灵敏度高、线性范围、精密度和准确度符合检测要求。 〔Objective〕 To seek a simple,accurate and sensitive method for detection of volatile phenolics in drinking water.〔Methods〕 Volatile phenolics in drinking water not been extracted by chloroform were detected by flow injection analysis (FLA).〔Results〕 The detection limit was 1.5μg/L and the linear range of standard curve was 1.5μg/L~50μg/L ( r ≥0.999).The relative standard deviation and recovery rate were 2.3%~7.4% and 88.9%~109%,respectively.〔Conclusions〕 The method is simple,sensitive,rapid and accurate.
出处 《中国卫生检验杂志》 CAS 2004年第3期300-301,共2页 Chinese Journal of Health Laboratory Technology
关键词 流动注射 饮用水 挥发性酚类化合物 测定方法 Flow injection analysis Drinking water Volatile phenolics
  • 相关文献

参考文献4

  • 1Written by Ninglan Liao Application Group,Quik Chem Method 10-210-00-3-A,2001.7.17.
  • 2Standard methods for the examination of water and wastewater.20th,method-phenols.
  • 3.生活饮用水水质卫生规范.[S].中华人民共和国卫生部卫生法制与监督司,2001.6.
  • 4.生活饮用水检验规范.[S].中华人民共和国卫生部卫生法制与监督司,2001.6.