直接输注电喷雾电离四极杆飞行时间质谱在十六种中药来源化合物定性分析中的应用(英文) 被引量:4

Application of Direct Infusion Electrospray Ionization Quadrupole Time-of-Flight Mass Spectrometry in Qualitative Analysis of Sixteen Compounds From Traditional Chinese Medicine
在线阅读 下载PDF
导出
摘要 采用直接输注电喷雾电离质谱对十六种中药来源的化合物进行了分析。观察了它们的电喷雾电离质谱和串联质谱特征 ,以了解不同结构化合物的电离能力和裂解特点。观察表明 :为使样品成分尽可能多地被检测 ,必须同时进行正负两种模式的分析。化合物结构中羰基的存在似有利于化合物在正离子模式下形成准分子离子或其它分子加和离子。当采样锥电压为 3 0 V时 ,大多数化合物能形成准分子离子或其它分子加和离子。但改变采样锥电压似不能完全避免源内裂解。建议分析时同时考察不同采样锥电压 (如 1 0 V、3 0 V和 5 0 V)。串联质谱分析被证明是获得进一步结构信息的有效手段 ,因为对于所有被观察的准分子离子和其它分子加和离子 ,均能在串联质谱分析中检测到碎片离子。本研究在四物汤的研究中已经成功地运用了该方法。 Direct infusion (DI) electrospray ionization quadrupole time-of-flight (Q-TOF) mass spectrometry was applied to analysis sixteen compounds from traditional Chinese medicines (TCMs). Characteristics of these compounds in electrospray ionization mass spectrometry (ESI-MS) and in MS/MS were observed to know the differences of ionization capacity and dissociation characteristics of structurally different compounds. The observation indicated that in order to maximize numbers of constituents detected, it is necessary to analyze sample in both positive and negative mode. The existence of carbonyl group in structure seems to be a factor that favors the formation of quasi-molecular ion (QMI) or other adduct molecular ion (OAMI) in positive mode. When cone voltage was 30V, most compounds could form mass spectral peaks of QMIs or OAMIs. However, in-source collision-induced dissociation (CID) seemed unable to be completely avoided by altering cone voltage. Analysis was recommended at three different cone voltages of 10V, 30V and 50V. MS/MS was proved to be an effective way to obtain additional structural information because fragment ions were observed for all QMIs and OAMIs in MS/MS. The developed method has been successfully applied in our study on Si-Wu-Tang. The method is useful for quick qualitative analysis of constituents in mixture samples of TCM.
出处 《质谱学报》 EI CAS CSCD 2004年第4期209-216,224,共9页 Journal of Chinese Mass Spectrometry Society
基金 Supported by the National Nature Science Foundation of China (30 2 716 17 30 0 70 913) and Beijing Science andTechnology Commission Project (H0 10 2 10 2 2 0 113)
关键词 中药 电喷雾电离 直接输注 飞行时间 定性分析 WT5”,6BZ]traditional Chinese medicine electrospray ionization direct infusion time-of-flight qualitative analysis
作者简介 LIANG Qian-de(1971-), Mail, Assistant Researcher, Major in Pharmaceutical Analysis&Modernization of Traditional Chinese Medicine. E-mail:lqd@nic.bmi.ac.cn,Corresponding author: WANG Sheng-qi(1962-),Mail, Researcher, Major in Nucleic Acid Drug&Molecular Diagnosis.E-mail: sqwang@nic.bmi.ac.cn
  • 相关文献

参考文献2

二级参考文献14

  • 1任仁安.中药鉴定学[M].上海:上海科学技术出版社,1992.87-88.
  • 2Cole R B. Some Tenets Pertaining to Electrospray Ionization Mass Spectrometry. J Mass Spectrom,2000,35(7):763.
  • 3Galand N, Pothier J, Dollet J, et al. OPLC and AMD, Recent Techniques of Planar Chromatography: Their Interest for Separation and Characterization of Extractive and Synthetic Compounds.Fitoterapia, 2002,73:121.
  • 4Masashi T, Shōko K, Mie ō. Water-Soluble Constituents of Rehmanniae Radix. Ⅰ. Carbohydrates and Acids of Rehmannia glutinosa f. fueichingensis. Chem Pharm Bull, 1971, 19(7) :1455.
  • 5Reinhold V N, Reinhold B B, Costello C E. Carbohydrate Molecular weight profiling, Sequence, Linkage, and Branching Data:ES-MS and CID. Anal Chem,1995,67(11):1772.
  • 6Masashi T, Shoko K, Mie o. Water-soluble constituents of rehmanniae radix. I. carbohydrates and acids of Rehmannia glutinosa f. Hueichingensis. Chem Pharm Bull, 1971,19(7): 1455.
  • 7Cole R B. Some tenets pertaining to electrospray ionization mass spectrometry . J Mass Spectrom , 2000,35(7):763.
  • 8Cooper H J, Marshall A G. Electrospray ionization fourier transform mass spectrometric analysis of wine. J Agric Food Chem, 2001,49(12) :5710.
  • 9姚新生.天然药物化学:第2版[M].北京:人民卫生出版社,1994.63.
  • 10何丽一.平面色谱方法及应用[M].北京:化学工业出版社,1999..

共引文献11

同被引文献67

引证文献4

二级引证文献39

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部