期刊文献+

血浆转氨酶在两种不同生化分析仪上的检测结果比较

Plasma Transaminase in Two Kinds of Different Biochemical Analyzer Results
在线阅读 下载PDF
导出
摘要 目的:通过应用不同的生化分析仪检测血浆转氨酶,来判定不同生化分析仪的准确性。方法选用美国 BECKMAN DXC600全自动生化分析仪和日本 HITACHI7080全自动生化分析仪,分别对400份患者的新鲜血浆,同时在两台仪器上进行血浆谷丙转氨酶(ALT)、碱性磷酸酶(ALP)、谷草转氨酶(AST)的测定。对两种全自动生化分析仪的测定结果间的可比性进行判断。结果两台不同全自动生化分析仪同时测出的结果数据,相似度高、数值很接近,差异不显著,无统计学意义P>0.05。结论不同型号的全自动生化分析仪,在对同一标本,并使用同一种方法情况下,得出的数据结果相似度很高,具有可比性,可以更好的服务于临床。 Objective To detect the plasma transaminase biochemical analyzer to determine the accuracy of dif erent applications,dif erent biochemical analyzer.Methods The America BECKMAN DXC600 automatic biochemical analyzer and the Japanese HITACHI7080 automatic biochemical analyzer,400 samples of fresh plasma of patients,and plasma alanine aminotransferase in two instruments(ALT),alkaline phosphatase(ALP),aspartate aminotransferase(AST)determination.The determination results of two kinds of automatic biochemical analyzer comparability judgement.Results Two dif erent automatic biochemical analyzer measured data at the same time,high similarity,numerical value is very close,the dif erence was not significant,no statistical significance of P>0.05.Conclusion Dif erent types of automatic biochemical analyzer,in the same sample,and use the same method of case,results from the similarity is very high,comparable,can bet er service to clinical.
作者 崔秀娟 CUI Xiu-juan(Department of Clinical Laboratory,Shenzhen Blood Center,Shenzhen 518035,Guangdong,China)
出处 《医学信息(医学与计算机应用)》 2014年第29期249-250,共2页 Medical Information
关键词 不同生化分析仪 血浆转氨酶 相似度 Dif erent biochemical analyzer Plasma transaminase Similarity
  • 相关文献

参考文献4

二级参考文献17

  • 1于嘉屏,潘柏申.血清酶学测定标准化进展[J].中华检验医学杂志,2005,28(12):1328-1330. 被引量:13
  • 2张秀明,庄俊华,徐宁,黄宪章,陈茶,尹一兵,梁伟雄.不同检测系统血清酶测定结果的偏倚评估与可比性研究[J].中华检验医学杂志,2006,29(4):346-349. 被引量:137
  • 3National Committee for Clinical Laboratory, Standards. Method comparison and bias estimation using patient samples[J].approved guideline, EP9-A, Pennsylvania, 1995, 15(7): 1-2.
  • 4Westgard J O , Bumett R W . Precision requirements for cost effection operation of analytical processes [J]. Clin Chem, 1990, (36) : 1629-1632.
  • 5Siekmann L, Bonora R, Bartis C A, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 degrees C. International Federation of Clinical Chemistry and Laboratory Medicine. Part 7. Certification of four reference materials for the determination of enzymatic activity of gamma-glutamyltransferase, lactate dehydro-genase, alanine aminotransferase and creatine kinase accord [J]. Clin Chem Lab Med. 2002.40 ( 7 ) : 739-745.
  • 6中华人民共和国国家标准,检测和校准实验室能力的通用要求[M].北京:中国标准出版社,2001:1-18.
  • 7The National Committee for Clinical Laboratory Standards.Method Comparison and Bias Estimation Using I'atient Samples Approved Guideline[S].EP9-A2.Pennsylvania,1995,15(7):1-2.
  • 8魏吴 丛玉隆.医学实验室质量管理与认可指南[M].北京:中国计量出版社,2004.27-80.
  • 9Moss DW.Enzyme reference materials:their place in diagnostic enzymology.J Int Fed Clin Chem,1994,6(1):4-6.
  • 10Lessinger JM,Ferard G,Grafmeyer D,et al.Usefulness of reference materials in calibration of enzyme activities.Eur J Clin Chem Clin Biochem,1995,33(11):859-864.

共引文献24

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部