摘要
根据JJF 1059.1-2012和JJG 1135-2017对热重分析仪(TGA)检定/校准过程中的测量不确定度进行了评定。结果表明:TGA检定/校准过程中带有差示扫描量热仪的TGA温度示值误差的扩展不确定度为0.70℃(k=2),单纯TGA温度示值误差的扩展不确定度为1.80℃(k=2),质量示值误差的扩展不确定度为12μg(k=2)。
According to JJF 1059.1-2012 and JJG 1135-2017,the measurement uncertainty in the verification/calibration process of thermogravimetric analyzer(TGA)were evaluated.The results show that the expanded uncertainty of temperature indication error of TGA with differential scanning calorimeter during TGA verification/calibration process was 0.70 ℃ (k=2).The expanded uncertainty of temperature indication error of TGA was 1.80℃ (k=2).The expanded uncertainty of mass indication error was 12μg(k=2).