基于Digital Diagnost型DR质量控制检测结果分析

Quality control analysis of test items based on Philips Digital Diagnost DR
原文传递
导出
摘要 目的探讨Digital Diagnost型DR质量控制检测结果,同时分析不同厂家DR的信号传递特性(STP)的差异,以期为检测人员的质控和行政人员日常监管提供参考。方法依据WS 76—2020《医用X射线诊断设备质量控制检测规范》,对Digital Diagnost型DR的AEC重复性、AEC响应、探测器剂量指示(DDI)、STP、高对比度分辨力、低对比度分辨力等检测项目进行质量控制检测,同时选取市场上常见DR分析其STP函数关系。结果Digital Diagnost型DR的DDI、高对比度分辨力和低对比度分辨力的基线值分别为影像中央像素值19733、2.8 Lp/mm和H5;其他检测项目质控结果均满足WS 76—2020的要求。常见DR的STP函数关系主要为线性和对数关系,但相同厂家不同型号DR也可能具有不同的STP函数关系。结论DR部分专用检测项目检测时,应注意多种因素的影响,同时细化检测项目的检测步骤与条件设置。在对STP进行分析时应按照厂家提供的函数关系进行模拟。 Objective To explore the quality control results of Digital Diagnost digital radiography(DR),and also to analyze the differences in signal transfer characteristics(STP)of DR from different manufacturers,in order to provide a reference for quality control of testers and daily supervision of administrative staff.Methods According to WS 76—2020“Specification for testing of quality control in medical X-ray diagnostic equipment”,the test items such as automatic exposure control(AEC)repeatability,AEC response,detector dose indication(DDI),STP,high-contrast resolution,and low-contrast resolution of Digital Diagnost DR were qualitatively controlled,and several common DRs on the market were selected to analyze their STP functional relationships.Results The baseline values of DDI,high-contrast resolution,and low-contrast resolution of Digital Diagnost DR were 19733,2.8 Lp/mm,and H5 of image central pixel value,respectively;the quality control results of other test items met the requirements of WS 76—2020.The STP functional relationships of common DRs were mainly linear and logarithmic,but different models of DR from the same manufacturer might also have different STP functional relationships.Conclusion When testing some special test items of DR,attention should be paid to the influence of a variety of factors,while refining the test steps and condition settings for the test items.The simulation should be performed in accordance with the functional relationship provided by the manufacturer when analyzing the STP.
作者 赵徵鑫 孙刚涛 翟自坡 王强 王海华 朱波 钱前 杨勇 翟贺争 ZHAO Zhixin;SUN Gangtao;ZHAI Zipo;WANG Qiang;WANG Haihua;ZHU Bo;QIAN Qian;YANG Yong;ZHAI Hezheng(Hangzhou Hospital for the Prevention and Treatment of Occupational Disease,Hangzhou 310014 China;Hubei Centers for Disease Control and Prevention,Wuhan 430070 China;Hunan Prevention and Treatment Institute for Occupational Diseases,Changsha 410021 China;Institute of Radiation Medicine,Chinese Academy of Medical Sciences,Tianjin 300192 China)
出处 《中国辐射卫生》 2022年第5期553-557,共5页 Chinese Journal of Radiological Health
基金 浙江省卫生健康科技计划(2021KY956,2022KY1043) 杭州市科技发展计划(20201203B220) 湖南省卫生计生委科研计划课题项目(C2017062)
关键词 DR 质量控制 信号传递特性 DR Quality control Signal transfer characteristics
作者简介 赵徵鑫(1988—),男,安徽亳州人,硕士,工程师,主要从事辐射检测与评价工作。E-mail:1005709146@qq.com;通信作者:翟贺争,E-mail:zhaihezheng@irm-cams.ac.cn
  • 相关文献

参考文献9

二级参考文献61

共引文献53

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部