摘要 移动式O型臂系统是主要应用于骨科领域的设备,如果移动式O型臂系统性能不可靠,则会严重影响医生对患者的诊断以及治疗,目前国内和各个省份都还没有相应的规程规范。通过对移动式O型臂系统主要组成及工作原理和检测用标准器的分析介绍,依据相关文献,同时考虑移动式O型臂系统检测的实际要求,该文确定6个项目作为检测项目,建立移动式O型臂系统检测方法。通过多次检测试验验证结果满足预期要求。论文证明了检测方法的可操作性及合理性。 The mobile O-arm systems is mainly used in orthopaedic field.If the performance of the mobile O-arm systems is not reliable,it will seriously affect the doctor's diagnosis and treatment of patients,at present,there are no relevant regulations and specifications in our country and every province.Based on the analysis and introduction of the main composition and working principle of the mobile O-arm systems and the standard instrument for detection,according to the relevant literature,at the same time,considering the practical requirements of mobile O-arm systems,this paper determines 6 items as detection items,and establishes a detection method of mobile O-arm system.The test results are verified to meet the expected requirements.The paper proves the operability and rationality of the detection method.
作者 李静 LI Jing(Henan Institute of Metrology and Testing Sciences,Zhengzhou 450000,China)
出处 《中国测试》 CAS 北大核心 2024年第S01期27-30,共4页 China Measurement & Test
关键词 移动式O型臂系统 辐射输出的空气比释动能率 辐射输出的质 3D透视 mobile O-arm system specific kinetic energy rate of the air emitted by radiation quality of the radiation output 3D perspective
作者简介 李静(1990-),女,河南舞阳县人,工程师,硕士,从事医学与电离辐射计量测试与研究。