双源CT前瞻性心电门控大螺距Flash扫描模式在支气管动脉CTA成像中的应用 被引量:2

Application of dual source CT prospective ECG gated large pitch flash scanning mode in bronchial artery CTA imaging
在线阅读 下载PDF
导出
摘要 目的探讨双源CT前瞻性心电门控大螺距扫描模式在支气管动脉CTA成像中的应用价值。方法选择2017年7月-2020年1月在浙江省丽水市中心医院用西门子第三代双源Force CT行支气管动脉CTA检查的患者90例,随机分为三组,每组30例,分别采用非心电门控螺旋扫描模式、非心电门控大螺距扫描模式及前瞻性心电门控大螺距扫描模式扫描,比较三组辐射剂量、主观及客观图像成像质量,客观评价图像质量包括胸主动脉CT值、噪声(SD)、信噪比(SNR)、对比噪声比(CNR)。结果辐射剂量方面,心电大螺距组DLP、CTDI_(vol)、ED低于非心电螺旋组及非心电大螺距组,差异均有统计学意义(P<0.05);非心电大螺距组DLP、CTDI_(vol)、ED低于非心电螺旋组,差异均有统计学意义(P<0.05)。心电大螺距组图像质量主观优良率高于非心电螺旋组及非心电大螺距组,差异均有统计学意义(P<0.05);非心电大螺距组图像质量主观优良率高于非心电螺旋组,差异有统计学意义(P<0.05)。三组图像主动脉CT值均能满足临床诊断要求,且三组主动脉图像CT值、SD值、SNR、CNR比较差异均无统计学意义(P>0.05)。结论在心率<90次/min的患者中,应用双源CT前瞻性心电门控大螺距Flash扫描模式进行支气管动脉CTA成像,图像质量与非心电门控下螺旋扫描和大螺距Flash扫描相比明显提高,且辐射剂量大幅度减少,仅为(0.66±0.07)mSv,具有较大的应用和推广价值。 Objective To evaluate the value of dual source CT large pitch prospective ECG gated scanning mode in bronchial artery CTA imaging.Methods 90 patients who underwent bronchial artery CTA with Siemens third-generation dual source force CT in Lishui Central Hospital of Zhejiang Province from July 2017 to January 2020 were randomly divided into three groups,with 30 cases in each group.They were scanned with non ECG gated spiral scanning mode,large pitch non ECG gated scanning mode and large pitch prospective ECG gated scanning mode respectively.The radiation doses of the three groups were compared Subjective and objective image quality.Objective evaluation of image quality includes thoracic aorta CT value,noise(SD),signal-to-noise ratio(SNR)and contrast noise ratio(CNR).Results In terms of radiation dose,DLP,CTDIvol and ed in ECG group were lower than those in non ECG spiral group and non ECG large pitch group(P<0.05);DLP,CTDIvol and ed in non ECG large pitch group were lower than those in non ECG spiral group(P<0.05).The subjective excellent and good rate of ECG large pitch group was higher than that of non ECG spiral group and non ECG large pitch group(P<0.05);The subjective excellent and good rate of image quality in non ECG large pitch group was higher than that in non ECG spiral group(P<0.05).The CT value of aorta in the three groups could meet the requirements of clinical diagnosis,and there was no significant difference in CT value(HU),SD value,SNR and CNR value among the three groups(P>0.05).Conclusion In patients with heart rate<90 beats/min,using dual source CT prospective ECG gated large pitch flash scanning mode for bronchial artery CTA imaging,the image quality is significantly improved compared with non ECG gated spiral scanning and large pitch flash scanning,and the radiation dose is significantly reduced,only(0.66±0.07)mSv,which has great application and popularization value.
作者 程淑芳 林桂涵 惠俊国 胡祥华 卢陈英 王海林 翁巧优 纪建松 陈春妙 CHENG Shufang;LIN Guihan;HUI Junguo;HU Xianghua;LU Chenying;WANG Hailin;WENG Qiaoyou;JI Jiansong;CHEN Chunmiao(Lishui Central Hospital,Lishui 323000,China)
机构地区 丽水市中心医院
出处 《浙江实用医学》 2021年第1期21-24,40,共5页 Zhejiang Practical Medicine
基金 浙江省医药卫生科技计划项目(2019KY808)
关键词 支气管动脉疾病 双源Force CT Flash扫描模式 心电门控 辐射剂量 bronchial artery disease dual source force CT flash scanning mode ECG gating radiation dosage
作者简介 通讯作者:纪建松,E-mail:jjstcty@sina.com
  • 相关文献

参考文献9

二级参考文献66

共引文献110

同被引文献20

引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部