Evaluation of Gamma Index Analysis for Detecting Errors in Patient-specific Quality Assurance in Intensity Modulated Radiotherapy Taylan Tugrul1(1.Department of Radiation Oncology,Medicine Faculty of Van Yüzü... Evaluation of Gamma Index Analysis for Detecting Errors in Patient-specific Quality Assurance in Intensity Modulated Radiotherapy Taylan Tugrul1(1.Department of Radiation Oncology,Medicine Faculty of Van YüzüncüYıl University,Van,Turkey)Abstract:Quality assurance practices performed before treatment are believed to identify various potential errors.In this study,2-dimensional(2D)dosimetric results were analyzed by making some intentional mistakes in six different treatment plans.In this way,the detectability of errors was investigated.In all segments of all treatment plans,one of the multileaf collimators was kept fixed at different positions on the central axis.In addition to multileaf collimators error,gantry error was also examined in the study.The dose distribution results obtained by Treatment Planning System(TPS)were compared with those obtained by the 2D array device,both as local calculation and global calculation methods,using the gamma analysis method.When the results are examined in the case where the Multi-leaf collimators(MLC)is fixed at the 1 cm position. 展开更多