The modified sum-peak method estimates radioactivity by using only the peak and the sum-peak count rates.To verify the modified sum-peak method,the dependence of the full energy peak efficiency on the source-to-detect...The modified sum-peak method estimates radioactivity by using only the peak and the sum-peak count rates.To verify the modified sum-peak method,the dependence of the full energy peak efficiency on the source-to-detector distance in a high-purity germanium detector system was studied using a Geant4 Monte Carlo simulation.The effect of the deadlayer in the germanium crystal was estimated by reference to experiments on 241 Am and the relative efficiency of the detector.The peak efficiency dependence on the source-to-detector distance was compared between the simulation and measurements.The modified sum-peak method is discussed with respect to these peak efficiencies.展开更多
文摘The modified sum-peak method estimates radioactivity by using only the peak and the sum-peak count rates.To verify the modified sum-peak method,the dependence of the full energy peak efficiency on the source-to-detector distance in a high-purity germanium detector system was studied using a Geant4 Monte Carlo simulation.The effect of the deadlayer in the germanium crystal was estimated by reference to experiments on 241 Am and the relative efficiency of the detector.The peak efficiency dependence on the source-to-detector distance was compared between the simulation and measurements.The modified sum-peak method is discussed with respect to these peak efficiencies.