浓缩同位素是校正质谱法、同位素稀释质谱法和双稀释剂法等同位素质谱分析技术的基础,其化学纯度及丰度量值的准确性直接影响着分析结果的准确性。但由于浓缩同位素中低丰度同位素的离子信号较难准确测量,且缺乏合适的标准物质校正,给...浓缩同位素是校正质谱法、同位素稀释质谱法和双稀释剂法等同位素质谱分析技术的基础,其化学纯度及丰度量值的准确性直接影响着分析结果的准确性。但由于浓缩同位素中低丰度同位素的离子信号较难准确测量,且缺乏合适的标准物质校正,给高精准的同位素丰度分析带来挑战。本文建立了适用于微量铕浓缩同位素样品纯化的锌还原-萃取色层法,纯化后的151 Eu和153 Eu两种浓缩同位素纯度优于99.99%,有效消除了其他稀土元素杂质的干扰。通过采用1012Ω信号放大器的法拉第杯接收样品中的低丰度同位素离子信号,建立了铕浓缩同位素的全蒸发-热电离分析方法。两种浓缩同位素样品的主丰度测量结果分别为151 Eu 0.9683676(11)和153 Eu 0.9876851(21),测量重复性比1011Ω信号放大器的法拉第杯以及文献中校正质谱法的测量结果提高了3倍。展开更多
The data of uranium isotope abundance ratios measured by thermal ionization mass spectrometer MAT-261 were analyzed in the three aspects including distribution function, correlation and uncertainty. The measurement da...The data of uranium isotope abundance ratios measured by thermal ionization mass spectrometer MAT-261 were analyzed in the three aspects including distribution function, correlation and uncertainty. The measurement data were confirmed to be normal distribution. Uncertainty and correlation of the isotope abundance ratios measured by MAT-261 in the certified uranium sample were investigated. The conclusion obtained indicated that, the correlation only between abundance ratios R5/8 and R4/8 is strong, but the others is weak for the enriched uranium sample; For the natural uranium sample, the correlation among all abundance ratios is weak. On this base, the uncertainty of all abundance ratios measured by MAT-261 had been evaluated.展开更多
文摘浓缩同位素是校正质谱法、同位素稀释质谱法和双稀释剂法等同位素质谱分析技术的基础,其化学纯度及丰度量值的准确性直接影响着分析结果的准确性。但由于浓缩同位素中低丰度同位素的离子信号较难准确测量,且缺乏合适的标准物质校正,给高精准的同位素丰度分析带来挑战。本文建立了适用于微量铕浓缩同位素样品纯化的锌还原-萃取色层法,纯化后的151 Eu和153 Eu两种浓缩同位素纯度优于99.99%,有效消除了其他稀土元素杂质的干扰。通过采用1012Ω信号放大器的法拉第杯接收样品中的低丰度同位素离子信号,建立了铕浓缩同位素的全蒸发-热电离分析方法。两种浓缩同位素样品的主丰度测量结果分别为151 Eu 0.9683676(11)和153 Eu 0.9876851(21),测量重复性比1011Ω信号放大器的法拉第杯以及文献中校正质谱法的测量结果提高了3倍。
文摘The data of uranium isotope abundance ratios measured by thermal ionization mass spectrometer MAT-261 were analyzed in the three aspects including distribution function, correlation and uncertainty. The measurement data were confirmed to be normal distribution. Uncertainty and correlation of the isotope abundance ratios measured by MAT-261 in the certified uranium sample were investigated. The conclusion obtained indicated that, the correlation only between abundance ratios R5/8 and R4/8 is strong, but the others is weak for the enriched uranium sample; For the natural uranium sample, the correlation among all abundance ratios is weak. On this base, the uncertainty of all abundance ratios measured by MAT-261 had been evaluated.