Investigating the minimum detectable activity concentration and contributing factors in airborne gamma-ray spectrometry 被引量:2
1
作者 Yi Gu Kun Sun +6 位作者 Liang-Quan Ge Yuan-Dong Li Qing-Xian Zhang Xuan Guan Wan-Chang Lai zhong-xiang lin Xiao-Zhong Han 《Nuclear Science and Techniques》 SCIE EI CAS CSCD 2021年第10期30-38,共9页
In this study,the theory of minimum detectable activity concentration(MDAC)for airborne gamma-ray spectrometry(AGS)was derived,and the relationship between the MDAC and the intrinsic effi-ciency of a scintillation cou... In this study,the theory of minimum detectable activity concentration(MDAC)for airborne gamma-ray spectrometry(AGS)was derived,and the relationship between the MDAC and the intrinsic effi-ciency of a scintillation counter,volume,and energy res-olution of scintillation crystals,and flight altitude of an aircraft was investigated.To verify this theory,experi-mental devices based on NaI and CeBr 3 scintillation counters were prepared,and the potassium,uranium,and thorium contents in calibration pads obtained via the stripping ratio method and theory were compared.The MDACs of AGS under different conditions were calculated and analyzed using the proposed theory and the Monte Carlo method.The relative errors found via a comparison of the experimental and theoretical results were less than 4%.The theory of MDAC can guide the work of AGS in probing areas with low radioactivity. 展开更多
关键词 Airborne gamma-ray spectrometry(AGS) Minimum detectable activity concentration(MDAC) Sensitivity